Table of Contents

Advertisement

Specifications

This section lists the principal performance specifications for the
HP G1999A (HP 5973 MSD with factory-installed CI.) These specifications are
subject to change without notice. These specifications do not apply to CI upgrades;
see the upgrade installation documents.
Sensitivity, Scan
Positive CI: 75:1 RMS for 1 µl of 100 pg/µl Benzophenone (m/z = 183.1)
Negative CI: 500:1 RMS for 1 µl of 1 pg/µl Octafluoronapthalene (m/z = 272.0)
Sensitivity, SIM
Positive CI: 10:1 RMS for 1 µl of 100 fg/µl Benzophenone (m/z = 183.1)
Negative CI: 10:1 RMS for 1 µl of 1 fg/µl Octafluoronapthalene (m/z = 272.0)
N O T E
Installation does not include demonstration of SIM sensitivity.
9