To Connect The Vacuum Gauge Controller - HP 5973 MSD Hardware Manual

Table of Contents

Advertisement

 Maintenance
7R FRQQHFW WKH YDFXXP JDXJH FRQWUROOHU

To connect the vacuum gauge controller

Materials needed:
Gauge controller (HP 59864B)
Triode gauge cable (8120-6573)
The high-vacuum gauge controller is required for operating the MSD in CI mode.
1HYHU FRQQHFW RU GLVFRQQHFW WKH FDEOH IURP WKH WULRGH JDXJH WXEH ZKLOH WKH 06' LV XQGHU
W A R N I N G
YDFXXP 5LVN RI LPSORVLRQ DQG LQMXU\ GXH WR EURNHQ JODVV H[LVWV
1HYHU FRQQHFW RU GLVFRQQHFW WKH FDEOH IURP WKH WULRGH JDXJH WXEH ZKLOH WKH FRQWUROOHU LV
W A R N I N G
WXUQHG RQ 'DQJHURXV YROWDJHV DUH SUHVHQW RQ WKH FDEOH
C A U T I O N
Do
1 Connect the triode gauge cable to the triode gauge tube.
2 Connect the other end of the triode gauge cable to the gauge controller.
3 Connect the power cord to the gauge controller.
4 Connect the other end of the power cord to an appropriate electrical
outlet.
5 Pump down the MSD.
100
QRW
use an HP 59864
(older model) triode gauge controller with the HP 5973 CI MSD.