 CI operation
7R SHUIRUP &, DXWRWXQH

To perform CI autotune

After the reagent gas flow is adjusted, the lenses and electronics of the MSD should
be tuned. Perfluoro-5,8-dimethyl-3,6,9-trioxidodecane (PFDTD) is used as the
calibrant. Instead of flooding the entire vacuum chamber, the PFDTD is introduced
directly into the ionization chamber through the GC/MSD interface by means of the
gas flow control module.
C A U T I O N
After the system has been switched from EI to CI mode, or vented for any other reason, the
MSD must be baked out for at least 2 hours before tuning. Longer bakeout is recommended
before running samples for optimal sensitivity.
There is a PCI autotune for methane only, as there are no PFDTD ions produced by
other gases in positive mode. PFDTD ions are visible in NCI for any reagent gas.
Always tune for methane PCI first regardless of which mode or reagent gas you wish
to use for your analysis.
Default (starting) CI tune parameters)
Reagent gas
Ion polarity
Positive
150 µΑ
Emission
Electron energy
150 eV
Filament
1
Repeller
3 V
Ion focus
130 V
Entrance lens (gain)
0 mV/amu
Entrance lens offset
20V
EMVolts
1200
Gas Off
Off
Gas select valve
A
Suggested flow
20%
Source temp
250°C
Quad temp
106°C
Interface temp
320°C
Autotune
Yes
56
Methane
Negative
Positive
50 µΑ
150 µΑ
150eV
150eV
1
1
3 V
3 V
130 V
130 V
0 mV/amu
0 mV/amu
20
20V
1200
1200
Off
Off
A
B
40%
20
150°C
250°C
106°C
106°C
280°C
320°C
Yes
No
Isobutane
Negative
Positive
50 µΑ
150 µΑ
150eV
150eV
1
1
3 V
3 V
130 V
130 V
0 mV/amu
0mV/amu
20
20V
1200
1200
Off
Off
B
B
40
20
150°C
250°C
106°C
106°C
280°C
320°C
Yes
No
Ammonia
EI
Negative
N/A
50 µΑ
35 µΑ
150eV
70
1
1 or 2
3 V
30 V
130 V
90 V
0 mV/amu
0 mV/amu
20
25
1200
1200
Off
On
B
N/A
40
N/A
150°C
230°C
106°C
150°C
280°C
280°C
Yes
Yes