Later it has been shown by Craig Neill (CSIRO) that this formula generally performs better with
respect to interpolation between calibration points and extrapolation outside the calibration
range. Based on this and recommendation from the Argo oxygen meeting in Brest 2011 the
above formula has been implemented in the Optode firmware (serial no. 1000 and above).
In order to use the "Stern-Volmer-Uchida" formula the property called Enable SVUformula must be set to
'yes'. The coefficients c
0
������������������������������������ ���� �������� ���� ���� ����[��������] = ��������������������������������
A possibility for linear correction of the O
For new Optodes (serial no. 1000 and above) the two-point calibration procedure (ref
Calibration Procedure chapter 6.2.2 ) will adjust the ConcCoef coefficients.
to c
are stored in the SVUFoilCoef property.
6
concentration was also introduced:
2
page 55 of 93
TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
+ ��������������������������������
[����
����
����
]'
����