Redreference, Calibration Coefficients And Salinity Compensation - Aanderaa 4330 Operating Manual

Oxygen optode
Table of Contents

Advertisement

Sensor Configuration
After installing the sensor turn power on and open the
Select the newly installed
in the lower part of the window.
Figure
3-2
System
Output settings.
The absolute oxygen level, µM, is a default parameter and cannot be disabled, refer TD262a.
To enable/disable a parameter:
Select the output parameter from the list, press
setting by clicking the box (box is now checked), press
We recommend that you enable all parameters in case of later use. The memory card storage capacity is
normally not a limitation for the SEAGUARD
coefficients and perform quality control on the data.

3.2 RedReference, Calibration Coefficients and Salinity Compensation

Special property settings and calibration coefficients are found in User Maintenance.
Open the
User Maintenance
and tap
Configure
in the lower part of the window. The user maintenance holds four submenus:
Node Description
Salinity setting
Calibration Coefficients
Enable RedReference
Oxygen Optode
which should appear in the list of sensors, and tap
The
System Configuration
that can be enabled/disabled by the user, refer Figure 3-2.
Enabled properties (Yes) are stored in the data logger:
Enable
Enable
Enable
Enable HumidityComp
of vapour pressure in the calculations of the output
parameters. Enable HumidityComp can be set to No if
measurements are performed in dry air conditions or if you
like to perform the humidity compensation as a post-
processing operation. Disabling this will typically lead to 1-3
% lower readings depending on the relative humidity of the
air.
Configuration;
View/Edit
®
. Raw data can be used e.g. to control calibration
from the
Menu
button. Select the
page 22 of 93
TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
System Configuration
holds a list of output parameters
Air Saturation in engineering units.
Raw data, refer chapter 2.2 .
Temperature in engineering units.
property activates compensation
in the lower part of the window, and change the
Save
to save and close the window.
Oxygen Optode
from the
Menu
button.
Configure
from the list of sensors,

Advertisement

Table of Contents
loading

Related Products for Aanderaa 4330

This manual is also suitable for:

48354831

Table of Contents