Set the
Enable AirSaturation, Enable
parameters in the output string.
Set the
Enable Text
to
(No)
to output the measured values without the descriptive text.
Set the
Enable Decimalformat
property to
(No)
to output the values in exponential format (higher resolution in the output parameter
value).
The
Interval
property describes the measurement interval in seconds; the Optode provides a set of
measured data for every measurement interval.
The
Salinity
property is set to 0 at the factory and the measurements must be post processed to
compensate for the salinity variations at the measurement site. If the salinity is known at the site and
relatively stable, you can set the Salinity property according to the known value and the measurements
will be salinity compensated inside the sensor before presentation (useful for Real-Time measurements).
The
Enable HumidityComp
the output parameters.
complete dry conditions (dry air) or if you like to perform the humidity compensation as a post-processing
operation. Measurements in dry conditions are more accurate when the
(No). The property is set to
When
Enable Redreference
set at the red reference (no fluorescence). The property can be set to
situations; contact AADI service department.
calibration; if the property is set to
Set the
Enable Sleep
to
stay continuously switched on between recordings (drains more power).
Temperature, and
(Yes)
for the optode to output a detailed text string with the measurements, or
to
(Yes)
to output the measured values in decimal format. Set the
property describes compensation of vapour pressure in the calculations of
Enable HumidityComp
(Yes)
at the factory.
is set to (Yes), the phase measurements are performed with a zero-point
Enable Redreference
(No)
the Optode must be recalibrated.
(Yes)
for the Optode to go to sleep between recordings, or
page 34 of 93
TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
Enable Rawdata
can be set to
(No)
if measurements are performed in
Enable HumidityComp
is set to
to
(Yes)
to include these
is set to
(No)
in special measurement
(Yes)
at the factory before
(No)
for the optode to