Figure 6-7 Set the Update Interval
Figure 6-8 Monitor Sensor Readings
Calibration
AADI has discovered that an after-curing occures in the foils during the first 1-3 months after the foil was
manufactured which typically leads to 1-4 % lower readings if the sensors are calibrated before the after-
curing has stopped. As of today methods are in place to cure all foils before they are mounted on the
Optodes and calibrated.
TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
The next window shows sensor information like the
Node Description, Product name and number, and
Serial number, refer Figure 6-7.
The number of sensor parameters and the processing time can be
viewed in the window.
Update Interval
Select an
Press
Start
to start monitoring the sensor readings.
The parameter reading in engineering units is shown as
illustrated in Figure 6-8. The reading updates according to the
update interval.
Press
Freeze
to temporarily stop the update; press
monitoring (Start is the same button as Freeze).
Press
Cancel
to stop monitoring.
The sensor readings should be according to the
description in chapter 6.2 .
page 41 of 93
for the sensor monitoring.
Start
to restart