6.2 Function Test
We recommend that you perform a function test of the sensor operating in air to verify the sensor
readings. Refer chapter 6.2.1 and chapter 6.2.2 , for a description of the function test procedure.
At sea level the oxygen saturation should be approximately 100% in air (90-110 %, depending on air
pressure, local oxygen production/consumption and if the temperature sensor measurements are
representative for the temperature of the foil). The saturation will be significantly lower when you breathe
on or near the sensing foil.
The measured temperature should also agree with the ambient temperature.
6.2.1 SEAGUARD
Leave the Oxygen Optode 4330/4835 mounted onto the SEAGUARD
refer TD262a and TD262b for operating instructions.
Figure 6-6 Select sensors to monitor
®
Applications
Open
Sensor Monitor
the function is mainly used for test purposes.
Select the Optode from the list and press Start, refer
Figure 6-6.
page 40 of 93
TD 269 OPERATING MANUAL
OXYGEN OPTODE 4330, 4831, 4835
June 2017
®
Platform. Power the instrument;
Administrative Tools-> Sensor Monitor
can be used as a direct reading of the sensor;