Replacing The Keypad - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

CARESCAPE Monitor B650

12.4.5 Replacing the Keypad

206
2062973-002
1. Pull the Trim Knob off.
2. Tear the Keypad panel from the Front Unit,
starting from the upper left corner.
NOTE: Try to leave as little glue as possible to the
front frame cover. If any glue remains clean the
surface with water-diluted alcohol.
Assembling the keypad:
-
Remove the release liner of the keypad.
-
Place the hole at right end of the keypad to
the Trim Knob shaft. Then carefully guide the
lower edge against the cover edging.
3. Tear the membrane keypad from the Front
Unit and detach the keypad connector that
connects it to the User Interface Board.
Assembling:
-
Clean the glue from the surface.
-
Remove the release liner from the membrane
keypad.
-
Connect the membrane keypad connector
first and direct the membrane into place.
-
Pat the membrane keypad carefully to stick it
to the front cover.
-
Assemble the Keypad.

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CARESCAPE Monitor B650 and is the answer not in the manual?

Table of Contents

Save PDF