Table 5. Fluke 27 II Calibration Adjustment Steps (cont.)
Function
(Switch Position)
(Microamps)
Table 6. Fluke 28 II Calibration Adjustment Steps
Function
(Switch Position)
(AC Volts)
(DC Volts)
(DC Millivolts)
(Ohms)
(Diode Test)
(Amps)
(Amps)
Adjustment Step
C-36
C-37
C-38
C-39
C-40
C-41
C-42
Adjustment Step
C-01
C-02
C-03
C-04
C-05
C-06
C-07
C-08
C-09
C-10
C-11
C-12
C-13
C-14
C-15
C-16
C-17
C-18
C-19
C-20
C-21
C-22
C-23
C-24
C-25
C-26
C-27
Input Value
0.0 μA ac, 0 Hz
30.0 μA, 60 Hz
60.0 μA. 60 Hz
600.0 μA, 60 Hz
6000.0 μA, 60 Hz
600.0 μA, 0 Hz
6000.0 μA, 0 Hz
Input Value
600.0 mV, 60 Hz
600.0 mV, 20 kHz
6.000 V, 60 Hz
6.000 V, 20 kHz
60.00 V, 60 Hz
60.00 V, 20 kHz
600.0 V, 60 Hz
600.0 V, 10 kHz
6.000 V
60.00 V
600.0 V
600.0 mV
60.00 mV
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
0.000 Ω
50.0 MΩ
3.000 V
6.000 A, 60 Hz
6.000 A dc
60.00 mA, 60 Hz
400.0 mA, 60 Hz
60.00 mA dc
400.0 mA dc
Digital Multimeters
Calibration Adjustments
21