Safety Information - Fluke 27 II Calibration Information Manual

Digital multimeters
Hide thumbs Also See for 27 II:
Table of Contents

Advertisement

27 II/28 II
Calibration Information

Safety Information

"Warning" and "Caution" Statements
A Warning identifies hazardous conditions and actions that could cause bodily harm or
death.
A Caution identifies conditions and actions that could damage the Meter, the equipment
under test, or cause permanent loss of data.
To prevent possible electrical shock or personal injury, follow
these guidelines:
2
Use this Meter only as specified in this manual or the
protection provided by the Meter might be impaired.
Do not use the Meter if it is damaged. Before you use the
Meter, inspect the case. Look for cracks or missing plastic.
Pay particular attention to the insulation surrounding the
connectors.
Make sure the battery door is closed and secured before
operating the Meter.
Replace the battery as soon as the battery indicator ()
appears.
Remove test leads from the Meter before opening the
battery door.
Inspect the test leads for damaged insulation or exposed
metal. Check the test leads for continuity. Replace
damaged test leads before you use the Meter.
Do not apply more than the rated voltage, as marked on the
Meter, between the terminals or between any terminal and
earth ground.
Never operate the Meter with the cover removed or the
case open.
Use caution when working with voltages above 30 V ac
rms, 42 V ac peak, or 60 V dc. These voltages pose a shock
hazard.
Use only the replacement fuses specified by the manual.
Use the proper terminals, function, and range for
measurements.
Avoid working alone.
When measuring current, turn off circuit power before
connecting the Meter in the circuit. Remember to place the
Meter in series with the circuit.