How To Replace The Batteries; How To Replace The Fuses - Fluke 27 II Calibration Information Manual

Digital multimeters
Hide thumbs Also See for 27 II:
Table of Contents

Advertisement

27 II/28 II
Calibration Information

How to Replace the Batteries

To avoid false readings, which could lead to possible electric
shock or personal injury, replace the battery as soon as the
battery indicator () appears. If the display shows "batt" the
Meter will not function until the battery is replaced.
Replace the battery as follows, refer to Figure 2:
1. Turn the rotary switch to OFF and remove the test leads from the terminals.
2. Remove the six Phillips-head screws from the case bottom and remove the battery
While lifting the battery door, ensure the rubber gasket stays attached to
the battery compartment barrier.
3. Remove the three batteries and replace all three with AA Alkaline batteries ().
4. Ensure the battery compartment gasket () is properly installed around the outside
5. Replace the battery door by aligning the battery compartment barrier with battery
6. Secure the door with the six Phillips-head screws.

How to Replace the Fuses

Referring to Figure 2, examine or replace the Meter's fuses as follows:
1. Turn the rotary switch to OFF and remove the test leads from the terminals.
2. Refer to step 2 under the How to Replace the Batteries section above to remove the
3. Remove the fuse compartment seal () from the fuse compartment.
4. Gently lift out the fuse compartment door () from the fuse compartment.
5. Remove the fuse by gently prying one end loose, then sliding the fuse out of its
6. Install ONLY specified replacement fuses with the amperage, voltage, and speed
7. Replace the fuse compartment door by aligning the arrow on the fuse door with the
8. Replace the fuse compartment seal by aligning the tab on the seal with the outline on
9. Refer to steps four through six under the Replacing the Batteries section above to
10
door ().
edge of the battery compartment barrier.
compartment.
battery door.
bracket ().
ratings shown in Table 7. The 440-mA fuse is shorter than the 10-A fuse. For correct
placement of each fuse, note the marking on the printed circuit board under each
fuse.
arrow on the case bottom and lowering the door into the fuse compartment.
the case bottom. Ensure the seal () is properly seated.
reinstall the battery door.
XWWarning
Note

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

28 ii

Table of Contents