AEMC FlexProbe User Manual

Flexible ac current probe

Advertisement

Quick Links

®
FlexProbe
Flexible AC Current Probe
USER MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for AEMC FlexProbe

  • Page 1 ® FlexProbe Flexible AC Current Probe USER MANUAL...
  • Page 2 If a malfunction occurs within the warranty period, you may return the instrument to us for repair or replacement free of charge, provided we ® have your REGISTRATION CARD on file. AEMC Instruments will, at its option, repair or replace the faulty material.
  • Page 3: Table Of Contents

    Features ....................5 Features ....................6 Ordering Information .................6 Accessories ....................6 Specifications ....................7 Instrument Compatibility................9 Operation....................9 ® Making Measurements with the FlexProbe ........9 Tips for Making Precise Measurements ..........9 Typical Response Curves..............11 Maintenance....................13 Warning ....................13 Cleaning....................13 Battery Replacement ................14 Repair and Calibration ................15...
  • Page 4: Warning

    • Read the safety specifications section prior to using the current probe. Never exceed the maximum voltage ratings given. ® • Safety is the responsibility of the operator. The FlexProbe must be used only by qualified personnel using applicable safety precautions. • ALWAYS de-energize the circuit before wrapping the ®...
  • Page 5: International Electrical Symbols

    Packaging ® Your FlexProbe consists of the following items: • Flexible probe with in-line electronic module • User manual • Two 3V lithium batteries (CR2032) •...
  • Page 6: Description

    FlexProbe - Flexible AC Current Probe Description ® The FlexProbe is a flexible AC current transformer composed of a flexible sensor and an electronic module. The flexible sensor permits measurements on conductors where standard clamp-on probes could not be used. In particular, it can be installed in tight spaces, around breaker panels, around cable bundles, around wide or large bus bars, or even wrapped around irregular shapes.
  • Page 7 ® FlexProbe Figure 1 1. Flexible Sensor (Ø 0.5”, 12.5mm) 5. Low Bat Indicator (red LED) ® 2. FlexProbe Connector/Latch 6. Range Selection Switch 3. Common: Black Banana Plug (-) 7. Electronic Module 4. Positive: Red Banana Plug (+) 8. Lead (6.5 ft, 2m)
  • Page 8: Features

    Includes a flexible probe with in-line electronic module, user manual, two 3V lithium batteries (CR2032), and a product warranty and registration card. Accessories Banana (Female) / BNC (Male) Adapter......Cat. #2118.46 ® For connection of FlexProbe to BNC terminals on scopes and other displaying instruments. - 6 -...
  • Page 9: Specifications

    ® FlexProbe - Flexible AC Current Probe Specifications ELECTRICAL Accuracy: 10mV/A: 5 - 180A; ± (1% of Reading + 500mA) 180 - 200A; ± (2% of Reading) 1mV/A: 10 - 1800A; ± (1% of Reading + 500mA) 1800 - 2000A; ± (2% of Reading)
  • Page 10 ® FlexProbe - Flexible AC Current Probe MECHANICAL SPECIFICATIONS Module Output: Two 4mm safety banana jacks Battery: Two 3V lithium batteries (CR2032) Battery Life: From 3.2V to 2.8V, 720 hrs typical (continuous use) Low Battery: Red LED blinks when battery voltage is low Dimensions (sensor): 24"...
  • Page 11: Instrument Compatibility

    AC voltmeter, or other voltage measuring instrument with an input impedance greater ® than 1MΩ. To achieve the best overall accuracy, use the FlexProbe with an AC voltmeter having an accuracy of 0.75% or better. Operation Please make sure that you have already read and fully understand the WARNING section on page 2.
  • Page 12 (Fig. 2). ® • The overall measurement accuracy is the sum of the FlexProbe accuracy and the displaying instrument accuracy. 100 A For best accuracy center the FlexProbe ®...
  • Page 13: Typical Response Curves

    ® FlexProbe - Flexible AC Current Probe Typical Response Curves: Typical Frequency Response (50A on 10mV/A Range) -0.5 1000 10000 Frequency (Hz) Typical Phase Shift (50A on 10mV/A Range) 1000 10000 Frequency (Hz) Typical Frequency Response (500A on 1mV/A Range) -0.5...
  • Page 14 ® FlexProbe - Flexible AC Current Probe Typical Response Curves (cont.): Typical Phase Shift (500A on 1mV/A Range) 1000 10000 Frequency (Hz) Current Derating Curve 2500 2000 1500 1000 1000 10000 Frequency (Hz) - 12 -...
  • Page 15: Maintenance

    • Avoid electrical shock: do not attempt to perform any servicing unless you are qualified to do so. ® • Do not perform any service while the Flexprobe is on any circuit. • Avoid electrical shock and/or damage to the instrument: do not get water or other foreign agents into the electronic module.
  • Page 16: Battery Replacement

    • If no signal can be measured and the low bat. indicator is not visible, check the batteries for extremely low levels. ® • Remove the FlexProbe from any circuit before replacing the battery. • To replace the battery, open rear case, replace battery and reassemble.
  • Page 17: Repair And Calibration

    N.I.S.T. (includes calibration certificate plus recorded calibration data). ® Chauvin Arnoux , Inc. ® d.b.a. AEMC Instruments 15 Faraday Drive Dover, NH 03820 USA Tel: (800) 945-2362 (Ext. 360) (603) 749-6434 (Ext.
  • Page 18 ® ® Chauvin Arnoux , Inc. d.b.a AEMC Instruments 15 Faraday Drive • Dover, NH 03820, USA www.aemc.com 99-MAN 100217 v6 08/18...

Table of Contents