Series Single-Phase Meter Testing; True Three Phase Meter Testing (Using Customer Load) - Radian RM-111 Operation Manual

Automated comparator
Table of Contents

Advertisement

SERIES SINGLE-PHASE METER TESTING

% Registration =
NOTE #2:
Number of elements corresponds to the number of meter elements in series.

TRUE THREE PHASE METER TESTING (USING CUSTOMER LOAD)

% Registration =
Reference Standard pulses B+C+D * Reference Standard Pulse Constant
NOTE #3:
If testing 3-phase, then 3 reference standards must be connected into ports B, C and D. If testing
2-phase, then 2 reference standards must be connected to ports B and C, B and D or C and D. If
two or three current inputs on the reference standards are connected in series, then the reference
pulse constant must be adjusted by the number of current inputs in series. See NOTE#1 for standard
testing.
88
100% * Meter Kh * Pulses per Rev. * #Revs. * #REF Current Inputs
Reference pulses B * Reference Standard Pulse * # of Elements
100% *Meter Kh * Pulses per Rev. * #Revs.
Calculations
Calculations
RM-111 Operations Manual