Radian RM-111 Operation Manual page 76

Automated comparator
Table of Contents

Advertisement

Application 10
Application 10
Application 10
Application 10
8. Set up the test board as if testing a meter. For this example, the variables will be set for a Form
1S meter: Kh=1.8, 120V, 10Amps, and one revolution.
9. Start the test on the test board. The standard must be on, and the test board's light detector
should now be waiting for the fi rst pulse to start the test.
10. After the test board's counter has counted the specifi ed number of pulses or disk revolutions it
stops the test.
11. The RM-111's display shows the Kh value entered in the test and the end count of the
standard(s). The test board itself shows the percent error, which is the percent error of the test
board.
75
RM-111 Operations Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents