System Description - Nuvair Pro O2 Operation Manual

Oxygen analyzers
Hide thumbs Also See for Pro O2:
Table of Contents

Advertisement

2.0 System Description

TM
The Pro O
and Pro O
2
levels in gases in the range of 0.1 - 100.0% O
application:
• Pro O
TM
– where pressurized gas must be analyzed and the pressure and flow of the
2
sample flow can be regulated
• Pro O
Remote
2
gas applications where the Analyzer must be located remote to the sensor
Pro O
Internal
Sensor
Flexible
Tubing
The Analyzer is a water and impact resistant unit compatible with outdoor and marine
environments. Pressurized gases must be regulated to one atmosphere absolute (0 P.S.I.) prior
to analyzing.
When used in breathing gas applications, redundant Analyzers must be used for verification. In
diving, for example, one Analyzer must be used to monitor oxygen during breathing gas
production and a second independent Analyzer must be used to verify the oxygen content of the
breathing gas prior to diver use.
When using the Analyzer for diving applications with mixed gases other than air, you
must first obtain proper instruction from a certified diving instructor with a nationally
recognized training agency qualified in mixed gas diving. Improper use of this analyzer
may result in incorrect gas analysis which can lead to serious personal injury or death.
NUVAIR
Pro O
Oxygen Analyzer Operation Manual
2
TM
Remote
Oxygen Analyzers ("the Analyzer") measure oxygen (O
2
TM
- where non-pressurized gas must be analyzed or for pressurized
TM
2
Flow
Adapter Cap
. The choice of Analyzer depends on the
2
Pro O
WARNING
5
Page
TM
Pro O
Oxygen Analyzer
2
TM
Remote
2
Flow Diverter
External Sensor
)
2
Detachable
6 ft. Electrical
Cord

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pro o2 remote

Table of Contents