Page 1
Operation Manual Pro O Pro O Remote Oxygen Analyzers 01.11 Andersen Medical Gas 12 Place Lafi fi tte Madisonville, LA 70447 http://www.TheMedicalGas.com 1-866-288-3783...
Page 2
Pro O Oxygen Analyzer If you have any questions on this equipment please contact Technical Support at: Nuvair 2949 West 5 Oxnard, CA 93030 Phone: 1-805-815-4044 FAX: 1-805-815-4196 Email: info@nuvair.com Hours: Monday through Friday 8:00 AM to 5:00 PM PST USA...
Indicates a potentially hazardous situation, which if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. NOTICE Notifies people of installation, operation or maintenance information which is important but not hazard-related. NUVAIR Page Pro O Oxygen Analyzer Operation Manual...
If the sealing device is torn or missing, contact your supplier for assistance. The sealing device must be removed prior to initial use and is not necessary to reuse. NUVAIR Page Pro O...
Direct Flow Method Sample Flow Method Flow is Directed to Flow is Delivered from Flow Orifice on Face Low Pressure Source of Flow Adapter Cap via Flexible Tubing Vent Hole NUVAIR Page Pro O Oxygen Analyzer Operation Manual...
Page 8
Insert Flow Diverter with o-ring into small bore of Adapter. • Press Fit Method – Remove the Flow Diverter o-ring and press fit the Diverter into a 9/16 inch diameter cylindrical bore in gas flow path. NUVAIR Page Pro O...
• Analysis of gases containing greater than 50% oxygen requiring calibration with pure oxygen or certified calibration gas • Analysis of gases at altitudes above sea level requiring correction for reduced atmospheric pressure, contact Nuvair for more information. NUVAIR Page...
Out Holes in Tubing to Gas Flow Adapter Sample Flow of 1-5 L/min Step 1 Step 2 Allow 15 Seconds for Display Reading to Stabilize Record Reading while Gas is Flowing Step 3 NUVAIR Page Pro O Oxygen Analyzer Operation Manual...
Page 13
Seconds for Valve & Remove Display Reading Analyzer to Stabilize Verify that Gas Record Reading Continues to Flow While Gas is from Valve. If Not, Flowing Repeat Procedure Step 3 Step 4 NUVAIR Page Pro O Oxygen Analyzer Operation Manual...
Seconds for Display Reading to Stabilize Record Reading while Gas is Flowing Gas Flow Pictured is the rear of a Nuvair Nitrox Generator with sensor placed in mixing tube. Step 1 Step 2 NUVAIR Page Pro O Oxygen Analyzer Operation Manual...
Screws Remove & Replace Old Battery Remove Front Cover Step 1 Step 2 Replace Front Cover Turn Analyzer - Do Not Pinch Wires Perform Air Reinstall Calibration Screws Step 3 Step 4 NUVAIR Page Pro O Oxygen Analyzer Operation Manual...
Page 16
Replace Thermistor Plug (Marked with a “T” on the plug” Once the Analyzer is put back together you need to turn on the Analyzer and Calibrate the Analyzer before operating the compressor. NUVAIR Page Pro O Oxygen Analyzer Operation Manual...
• Eye Contact - Flush eyes with clean, fresh water for at least 15 minutes and get medical help immediately. • Skin Contact - Flush the affected area with clean, fresh water for at least 15 minutes and removed contaminated clothing. If stinging persists get medical attention. NUVAIR Page Pro O Oxygen Analyzer Operation Manual...
Page 18
Electrical Connector Note: Reversing Polarity Will Cause Display to Read Negative Step 5 Replace Turn Analyzer Front Cover - Do Not Pinch Wires Perform Air Reinstall Calibration Screws Step 6 Step 7 NUVAIR Page Pro O Oxygen Analyzer Operation Manual...
Page 19
Pro O Oxygen Analyzer The following pictures illustrate the steps required to replace the Sensor in the Pro O Remote when Installed on a Nuvair Nitrox Generator or Nitrogen Generator. Pro O Remote Disconnect Remove and Electrical Replace Old Cord from...
Restrictor A variety of Flow Restrictors and Pressure Regulators Regulator for the Sample Flow Method are available from Nuvair, all calibrated to produce a flow rate of 1 – 5 L/min with a Regulator output of 100 – 160 P.S.I.
Sensor not in air Check flow adapter High altitude Use altitude calibration procedure Reading drifts Rapid temperature change Stabilize temperature & recalibrate *Contact Nuvair for more information. Analyzer Specifications Range: 0.1-100.0% Oxygen (0-1 ATA PPO Display Accuracy: +/- 0.1% Sensor Type:...
Note that the calibration correction value is never used when the temperature and humidity conditions of the gas being measured are the same as the conditions during calibration. NUVAIR Page Pro O Oxygen Analyzer Operation Manual...
NUVAIR has been advised of the possibility thereof. NUVAIR makes no other warranty or representation of any kind, except that of title, and all other warranties, express or implied, including warranties of merchantability and fitness for a particular purpose, are hereby expressly disclaimed.
Page 24
Pro O Oxygen Analyzer Nuvair™ 2011 All Rights Reserved Phone (805) 815-4044 Fax (805) 815-4196 2949 West 5th Street Oxnard, CA 93030 Email : info@nuvair.com Web : www.nuvair.com Revision 0111...