Nuvair Pro CO Alarm User Manual

Carbon monoxide analyzer
Hide thumbs Also See for Pro CO Alarm:

Advertisement

Quick Links

User Manual
Pro CO Alarm
Carbon Monoxide Analyzer
Rev. 03.19

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Pro CO Alarm and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Nuvair Pro CO Alarm

  • Page 1 User Manual Pro CO Alarm Carbon Monoxide Analyzer Rev. 03.19...
  • Page 2 Gas flow to the analyzer should be regulated to .5-3 L/min with no pressure on the sensor. Pressurizing the sensor will lead to inaccurate readings and could damage it. If using a Nuvair 9517.6 flow reducer for sample flow, the incoming pressure should be 75-100 PSI.
  • Page 3 Pro CO Carbon Monoxide Analyzer If you have any questions on this equipment please contact Technical Support at: Nuvair 1600 Beacon Place Oxnard, CA 93033 Phone: +1 805 815 4044 FAX: +1 805 486 0900 Email: info@nuvair.com Hours: Monday through Friday...
  • Page 4: Table Of Contents

    6.0 Powering Off……………………………………………………………………………………. 7.0 Factory Reset…………………………………………………………………………………... 8.0 Maintenance……………………………………………………………………………………. 8.1 Analyzer Care…………………………………………………………………………… 8.2 9V Battery Replacement…………………………………………………………….… 8.3 Sensor Replacement……………………………………………………………..……. 8.3.1 Handling Sensor………………………………………………………..…………… 9.0 Spares and Accessories……………………………………………………………………….. 9.1 Sensor………………………………………………………………………………..…. 9.2 Calibration Equipment………………………………………………………………….. 10.0 Troubleshooting………………………………………………………………………….…… 11.0 Relay Output Schematics…………………………………………………………………… Appendix Analyzer Specifications……………………………………………………………………………. Warranty……………………………………………………………………………………………… www.nuvair.com...
  • Page 5: Introduction

    Indicates a potentially hazardous situation, which if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. Notifies people of installation, operation or maintenance information which is important but not hazard-related. www.nuvair.com...
  • Page 6: System Description

    Extreme CO exposure levels directed at the Analyzer sensor may damage the sensor. Don’t test the sensor in the direct flow of any engine mufflers/exhausts or any other known high concentrations of CO. www.nuvair.com...
  • Page 7: Controls

    Analyzer and is user-replaceable. The battery should be removed any time the Analyzer will be stored without use for extended periods of time. Screen will blink alternately from “000” to “bAt” at start up when battery is low. Low Battery Indication www.nuvair.com...
  • Page 8: Flow Adapter Cap

    CO calibration gas with a 10 ppm concentration and flow rate of 1 L/min. The equipment to produce this flow is available from Nuvair. See Spares and Accessories section. To assure the greatest accuracy for other applications, use the calibration gas concentration closest to the expected concentration in the gas being measured.
  • Page 9: Sample Flow Method

    3. Press the On/Off and Prog buttons simultaneously and hold until the screen flashes “Cal” 4. When the display returns to reading 000, the zero has been set. Nuvair offers CO gas testing kits - see addendum for more information 3.1. Sample Flow Method of Checking Calibration (Preferred) Step 1. Attach Step 2.
  • Page 10: Operation

    Never expose the sensor to pressures above atmospheric pressure (0 psi) or you may cause damage to the sensor and/or receive false readings. Damaged Sensors will not provide accurate gas analysis. Inaccurate gas analysis can lead to serious personal injury or death. www.nuvair.com...
  • Page 11: Programming Procedures

    • If analyzing a gas flow, the Sample Flow Method is the preferred method. The flow rate must equal .5 to 3 L/min at atmospheric pressure. To produce this flow, a Flow Restrictor and Regulator may be required. Contact Nuvair if you need assistance. 4.1. Programming Procedures Keep the Prog button pressed for more than two seconds and then release the button.
  • Page 12: Alarm Setting

    6.) To complete your entry and save the FSC value, press the On/Off button repeatedly until the next desired program page is displayed or the end screen appears. 7.) This procedure is used ONLY when a new sensor has been ordered from Nuvair and is ready to be installed. www.nuvair.com...
  • Page 13: Conversion Value Of Carbon Monoxide Sensor

    7.) This procedure is used ONLY when a new sensor has been ordered from Nuvair and is ready to be installed. The conversion value of the Carbon Monoxide sensor is set in factory and must be changed only when the CO sensor is replaced.
  • Page 14: Gain Factor

    7.) This procedure is used ONLY when a new sensor has been ordered from Nuvair and is ready to be installed. Then gain factor is set in factory and must be changed only when the CO sensor is replaced. The new sensor will come from the factory with a label showing the new ‘Fct’...
  • Page 15: Calibration Value

    Pro Co Analyzer will remain in alarm mode until the analyzed value goes below the alarm set point. The relay output typically is used to shut down the compressor. Nuvair can supply the necessary components to adapt your compressor to the relay or provide them at the time of install on a new compressor.
  • Page 16: Factory Reset

    • Do not immerse in liquid, leave unprotected outside, or store in a wet environment. • Protect Analyzer from excessive shock and impact. • Protect Analyzer from excessive exposure to sunlight and extreme temperatures. • Do not use the Analyzer in a hyperbaric environment. www.nuvair.com...
  • Page 17: Battery Replacement

    Back Cover Step 4. Replace Back Cover - Step 6. Turn Do Not Pinch Analyzer On Wires Step 7. Step 5. Perform Reinstall Calibration Screws *Contact Nuvair for replacement if the battery life of a Lithium battery should noticeably degrade. www.nuvair.com...
  • Page 18: Sensor Replacement

    • Eye Contact - Flush eyes with clean, fresh water for at least 15 minutes and get medical help immediately. • Skin Contact - Flush the affected area with clean, fresh water for at least 15 minutes and removed contaminated clothing. If stinging persists get medical attention. www.nuvair.com...
  • Page 19 Connector & Reconnect Note: Reversing Polarity Will Cause Display to Read Negative Step 8. Step 10. Replace Replace Flow Adapter Cap Front Cover - Do Not Pinch Wires Step 11. Turn Analyzer On Step 9. Reinstall Step 12. Check Screws Calibration www.nuvair.com...
  • Page 20: Spares And Accessories

    1L/min Flow delivered at a specific flow rate and pressure. Restrictor/ Regulator A variety of calibration gas canisters are available Assembly from Nuvair, with compatible Flow Restrictor/ Regulator assemblies to regulate the gas. 10 PPM CO Calibration Gas Canister 10.0 Troubleshooting...
  • Page 21: Relay Output Schematics

    11.0 Relay Output Schematics 4-20mA Connections: The relay output uses a special “Mini Jack” plug that can be purchased separately from Nuvair. Insert the plug into the output jack. Connections diagram below for additional wiring. The plug should be connected or disconnected when the instrument is switched off, or the instrument will automatically switch off.
  • Page 22: Appendix

    14 to 122°F (-10 to 50°C) Storage Temperature: 14 to 140°F (-10 to 60°C) Operating Pressure: Not to exceed 1 atmosphere absolute (0 psi) Operating Humidity: Non-condensing: 15-90% continuous 0-99% intermittent Note: All specifications are at ambient / sea level, 77°F / 25°C www.nuvair.com...
  • Page 23: Warranty

    Nuvair. Terms: Nuvair warrants the Pro CO to be free from defects in material and workmanship for a period of twelve (12) months from date of purchase. The warranty covers parts and labor. A warranty registration card, supplied with system documentation, must be filled out and submitted to Nuvair for the warranty to be registered.
  • Page 24 Nuvair Phone +1 805 815 4044 Fax +1 805 486 0900 1600 Beacon Place Oxnard, CA 93033 info@nuvair.com www.nuvair.com Rev: 03.19...