Tee Adapter; Troubleshooting - Nuvair Pro O2 Operation Manual

Oxygen analyzers
Hide thumbs Also See for Pro O2:
Table of Contents

Advertisement

6.3. Tee Adapter (Pro O
When using the Pro O
gas flow, the Tee Adapter is used to install the Sensor
and Flow Diverter in the gas path.

7.0 Troubleshooting

SYMPTOM
Battery symbol
No display
Zero reading
Reading erratic
Reading does not change when
calibration knob is turned
Display segments missing
Will not calibrate
Reading drifts
*Contact Nuvair for more information.
Analyzer Specifications
Range:
Display Accuracy:
Sensor Type:
Expected Sensor Life, Room Air:
Power:
Response Time:
Value
Operating Temperature:
Storage Temperature:
Pressure:
Humidity:
Warranty:
Note: All specifications are at ambient / sea level, 25°C
NUVAIR
Pro O
Oxygen Analyzer Operation Manual
2
TM
Remote
)
2
TM
Remote
to analyze in-line
2
Low Battery
Switched off
Bad connection
Sensor disconnected
Sensor expired
Pressure on sensor
Radio transmission
Sensor old or faulty
Condensation on sensor.
Faulty connections
Sensor failure
Display faulty
Sensor faulty
Sensor not in air
High altitude
Rapid temperature change
Tee Adaptor
Part No. 6024
REASON
Change the battery
Switch on
Check display connection
Check battery connection
Check connection
Change sensor
Check flow
Move unit away
Change sensor
Dry in air
Check connections
Change sensor
Return to dealer
Change sensor
Check flow adapter
Use altitude calibration procedure
Stabilize temperature & recalibrate
0.1-100.0% Oxygen (0-1 ATA PPO
+/- 0.1%
Electrochemical
36 Months
9V Alkaline Battery
Less Than 6 Seconds to 90% of Final
32-104°F (0-40°C)
32-122°F (0-50°C)
Sensitive to Partial Pressure
0-99% RH (Non-Condensing)
36 Months Pro-Rated
21
Page
TM
Pro O
Oxygen Analyzer
2
SOLUTION
*
)
2