4.0 Operation

Prior to each Analyzer use:
1) Verify that Sensor sealing device is removed
2) Turn unit on and monitor Display for low battery warning
3) Calibrate Analyzer as required.
Never expose the Oxygen Sensor to pressure or you may cause damage and/or false
readings. Damaged Sensors will not provide accurate gas analysis. Inaccurate gas
analysis can lead to serious personal injury or death.
TM

4.1. Pro O

2
TM
The Pro O
can be used to analyze a regulated gas sample flow, the contents of a gas
2
cylinder, or the flow from a regulator. The flow rate of gas must equal 1 – 5 L/min. To produce
this flow, a Flow Restrictor and Regulator may be required. See Spares and Accessories
section.
Gas, even under moderate pressures, can cause extreme bodily harm. Never allow any
gas stream to be directed at any part of your body.
Attach Flexible
Tubing to Gas
Sample Flow
of 1-5 L/min
Allow 15 Seconds for
Display Reading to
Record Reading
while Gas is Flowing
NUVAIR
Pro O
Oxygen Analyzer Operation Manual
2
WARNING
WARNING
Sample Flow Method
Step 1
Stabilize
Pro O
Verify that
Gas is Flowing
Out Holes in
Flow Adapter
Cap
Step 3
12
Page
TM
Oxygen Analyzer
2
Step 2

Advertisement