MasterCraft 052-0060-2 Instruction Manual page 4

Pocket digital multimeter
Hide thumbs Also See for 052-0060-2:
Table of Contents

Advertisement

3
model no. 052-0060-2 | contact us 1-800-689-9928
When servicing the meter, use only specified replacement
parts.
Use caution when working with voltage above 30 V RMS, 42 V
peak, or 60 V DC. Such voltages pose a shock hazard.
When using the probes, keep your fingers behind the finger
guards on the probes.
Connect the common test lead before you connect the live test
lead. When you disconnect test leads, disconnect the live test
lead first.
Remove the test leads from the meter before you open the
battery cover or the case.
Do not operate the meter with the battery cover or portions of
the case removed or loosened.
To avoid false readings, which could lead to possible electric
shock or personal injury, replace the battery as soon as the
battery indicator (
To avoid electrical shock, do not touch any conductor with hand
or skin.
Do not operate this instrument if your hand, a test lead or the
instrument is wet.
Remaining endangerment:
When an input terminal is connected to dangerous live potential
it is to be noted that this potential can occur at all other terminals!
CAT III - Measurement Category III is for measurements performed in
building installation. Examples are measurements on distribution
boards, circuit breakers, wiring, including cables, bus-bars, junction
boxes, switches, socket-outlets in the fixed installation, and equipment
for industrial use and some other equipment, for example, stationary
motors with permanent connection to the fixed installation.
Do not use the meter for measurements within Measurement Categories
IV.
) appears.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents