C-SMART-IPro
9.2 Cleaning,Disinfection and Sterilization
(according to DIN EN ISO 17664)
The procedure for cleaning, disinfection and Sterilization applies only to the
accessories lip hook, file clip and contra-angle.
9. 2. 1 Pre-treatment
Pulp and dentin residues must be removed immediately from the accessories
(within maximum 2 hrs). Do not let them dry!After the accessories have been
used on patients, place the accessories for cleaning,pre-disinfection and
interim storage directly into a bowl filled with an appropriate cleaning and
disinfecting solution (for max. 2 hrs). Then clean the accessories under
running water or clean in a disinfecting solution to remove all visible
contamination. The disinfectant should be aldehyde-free (aldehyde fixes
blood stains),tested for effectiveness (e.g. CE mark), suitable for
accessories disinfection and compatible with the accessories. Only
use clean, soft brushes to manually remove contamination or a clean,
soft cloth which you only use for this purpose. Do not use metal brushes or
steelwool. For better cleaning of the inner parts, the file clip must be pressed
and released five times during cleaning process. Please note that
disinfectants used for pre-treatment are only for personal protection and do
not replace disinfection when cleaning is completed. The pre-treatment
process should be performed in every case.
.&.
CAUTION
Do not use an automated procedure or ultrasonic bath to clean or disinfect
the accessories.
9.2.2 ManualCleaning and Disinfection
Cleaning
• Place the pre-cleaned accessories into the cleaning bath for the prescribed
contact time, the accessories must be sufficiently covered (if necessary careful
brushing with a soft brush). For better cleaning of the inner parts,the file clip
must be pressed and released five times during cleaning process.
·Then remove the instruments from the cleaning bath and rinse them thoroughly
with water for at least three times for 1 min. and press and release the file clip
five times.
Disinfection
·Place the cleaned and inspected accessories into the disinfection bath for
the prescribed contact time; the accessories must be sufficiently covered.
For better disinfection of the inner parts, the file clip must be pressed and
released five times during disinfection process.
U11er Manual
C-SMART·IPro
Lt.
WARNINGS
• Do not disassemble or alter the motor handpiece.
• Do not expose the unit,motor handpiece or battery charger to any liquid.
• Do not expose the device to direct or indirect sources of heat. Operate and
store the device in a safe environment.
Do not drop the device.
• In order to avoid possible risks due to electromagnetic
use any electrical medical device or electrical device of any other kind in
close proximity to the device.
• The device may possibly malfunction if used in the presence of an
electromagnetic interference wave. Do not install the device in the vicinity
of any device that emits magnetic waves.
• Do not use the device in the presence of free oxygen or anesthetic
substances or flammable products.
• None of the device components are delivered disinfected or sterilized:
components such as control unit,motor and motor cable need to be
disinfected,the contra-angle and Lip hook needs to be sterilized prior to
first use and in between each patient!
• The plastic enclosure is not sealed, do not use any liquid or spray directly
on the unit, especially on the monitor or near the electrical sockets.
• Follow the file manufacturer's instructions for use of the endodontic files.
• The file system shown on the display must always match the file in use. This
is of the utmost importance in order to avoid misusing reciprocating files and
continuous rotary files.
Uoer Manual
interference,
do not
II