Coxo C-SMART-1 Pro User Manual page 25

Endodontic endo motor apex locator
Table of Contents

Advertisement

C·SMART-l Pro
0.0
Ml
,f.. •.,
Fig.22 ApexArea-Red
(7'
NOTE
The apex indication line shows the file tip position inside the root canal:
• Blue section:
WARNING section, close to the apex area.
• Green to yellow section:
Root apex area
• Red section:
Apex foramen is passed
,&caution
As with all electronic length determination devices, the bars shown in the root
apex area do not represent the specific size.
Interruption of measurement
During the length measurement the File clip may be disconnected from the
file and reconnected at any time (e.g.when the file is changed to a larger
size or when the length of another root canal should be measured). The
device detects automatically that a new length measurement cycle is
initiated.
,
Uaar Manual
\(_
Ml
<· •.,
Fig.23 Beyond the apex foramen
-Red dot
C-SMART-lPro
Displays the AUTO REVERSE function
(disabled for reciprocating systems).
AUTO
AUTO REVERSE on
7
REVERSE
-
AUTO REVERSE off:
•11
See8.3.2.6 Auto Reverse
E;jl
Means the motor handpiece is connected to the
Control unit and LED lighting function is not
available.
Motor and
8
LED
Means the motor handpiece is connected to the
Lighting
Control unit , LED lighting function is available
and the delay time is 3S.
See8.3.2.6 LED lighting function
8.3.2.4 System and File Select
File Library
The device contains a file library, the preset NiTi systems please refer the
The manufacturer reserves the right to update the file library
torque card.
and the systems contained in it.
The device contains a user-created system called PROGRAM. There are 7
continuous rotary files
,&caution
• Follow the file manufacturer's instructions for use of endodontic files.
·The file system shown on the display must always match the file in use. This
is of the most importance in order to avoid misusing reciprocating files and
continuous rotary files.
·Torque and speed values are subject to change by the file manufacturers
without notice. Therefore, the preset values in the library must be checked
prior to use.
• Torque values shown on the display are accurate and reliable only with
contra angle properly maintained and lubricated.
Uoar Manual
The motor handpiece is not connected to
the Control unit

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the C-SMART-1 Pro and is the answer not in the manual?

Questions and answers

Table of Contents

Save PDF