C-SMART-l Pro
8-3.1.3
working
1) Starting Length Measurement
• Connect the File clip and the lip hook to the test wire.
• Place the lip hook on the patient's lip on the opposite side of the tooth to be
treated.
• Insert the file into the root canal and hold the file clip in the metal part of the
file (Fig.
17)
• The file movement in the root canal is shown on the left side of the screen
work area (Fig.
16)
r:::J='
NOTE
If no progress of the file indicate faulty connection:
• Check proper connection of the test wire.
• Clean the File clip contact.
• Irrigate the canal,if necessary, and start again.
&.
CAUTIONS
• We recommend that not to continue the measurement if the faulty
connection appears.
• Go to the CHECK Mode to check the performance of the device
8.4).
(see chapter
Measurement
2)
Coronal and Medial Area
• Slowly insert the measuring file into the canal.
• File movement along the coronaland medial section towards the apex area
is represented on the full root canal image by the file continuously moving
down
(Fig.18).
Uaar Manual
Fig.17
C-SMART·lPro
&.
CAUTION
Before using the motor handpiece, verify the correctness of the changed
parameters.
8.3.2.6
Auto Reverse
r:::J='
NOTE
In reciprocating motion,the auto reverse function is disabled.
AUTO REVERSE on:
If, during operation the load reaches the preset torque limit value, the motor
handpiece will automatically rotate in the reverse direction. When the load is
removed, the motor handpiece returns to normal forward rotation automatically.
,T.,
-+
I
Further Loading
Load within the
beyond the preset
Torque Limit Value
Torque Limit Value
I
I
Forward Rotation!
Reverse Rotation
AUTO REVERSE off:
If, during operation the load reaches the preset torque limit value, the motor
handpiece will stop without reverse rotation.
If you want the file to rotate forward again,press the ON/OFF key once.
On/off Auto Reverse Function:
Uoer Manual
f
-+
I
When the load is
reduced, The motor is
automatically restored
I
I
Forward
Rotation!
0
SETUP
A
English
.,.
ectdlfl'erelltllnguegu
:! fT ;.AnrwWMmode
A
ON
•
F
:
!!t
g bl
ON
•
'!.'!tL
b dll.ytlm..
,.-
&
:
-
eg
8
JIIrlltiiNoMIIMry
MI'III'.f
B
Need help?
Do you have a question about the C-SMART-1 Pro and is the answer not in the manual?
Questions and answers