7.5 Charging

Charge the battery according to the following steps:
• Connect the charger's connector to the Control unit's charging socket
G)
(Fig.12).
• Connect the plug of the charger to a power outlet
• It takes about 4 hours to fully charge and about 5 hours for fully discharged
batteries (E.g., the device is not used for a long time).
• When charging, the charging status will be displayed on the screen. (Fig.13)
Fig.12
r:::T
NOTE
In case of difficulty in inserting the connector, the polarity may be incorrect
Do not insert it by force.
-
7
0%
Fig.13
C-SMART-1 Pro User Manual
8.4 Functional check of the apex locator
It is recommended to check the performance of the Apex Locator once a week.
With the built-in inspection function, the special test plug can automatically
check the basic functions of the instrument in the first step, and then inspect the
accessories in the second step (Fig. 29).
e
0
0
0
Fig.29 Device inspection
The specific steps to use the check function , as follows:
• Disconnect the test wire, adaptor and motor handpiece from the device.
·Device CHECK will start automatically if the tester inserted in the device and
the results will be shown on the screen.
• Display shows OK (Fig. 30) ,Indicates that the device function is
·Display shows ERROR (Fig_ 31),Indicates that there is a fault
0
0
Fig.30 Device is OK
.&.
CAUTION
ERROR message indicates that the device is not operating properly.
Please contact your local dealer or contact the factory directly for assistance_
• Disconnect the tester from the device and prepare the test of the wire with
accessories.
normal.
0
0
Fig.31 Device is ERROR