Calibrating The Analyzer; Start Calibration - ABB Navigator 540 Instruction Manual

Measurement made easy
Hide thumbs Also See for Navigator 540:
Table of Contents

Advertisement

6

Calibrating the analyzer

Quick calibrations are initiated via the CAL prompt displayed on
Operator pages. Scheduled calibrations and other calibration
routines must run from the Calibrate level menus (accessed only via
the Advanced access level – see Step
Caution.
Do not calibrate the analyzer until the wet-section and
transmitter are installed and ready for operation.
Allow at least 1 hour for sodium / hydrazine sensors to
stabilize before running a calibration.
Calibrating for the first time
Quick calibration routines (low level dissolved oxygen, hydrazine,
sodium) differ only in the calibration type that is performed:
LLDO = air cal.
hydrazine = high cal.
sodium = two-point cal.
To perform a quick calibration from an Operator page:
1. Press the
key:
Navigator
Tag 1
4.
4.2
21.7
21.7
LLDO
The Calibrate page is displayed:
Menu
Calibrate
Exit
2. Press the
key – the Calibrate / Sensor 1 (2, 3, 4) page is
displayed with all available wet-sections shown:
Calibrate
Sensor 1
Exit
Use the
/
keys to select the wet-section to
be calibrated.
10
7
.
10 : 34: 52
2012–04–15
ppb
°C
CAL
Select
Select
3. Press the
key – the Sensor 1 (2, 3 or 4) / Start Calibration
page is displayed:
Sensor 1

Start Calibration

Exit
4. Press the
key – the Calibration page is displayed with a
bar graph indicating calibration progress:
Low Level D. O. 1
PV 95.0
Please Wait
Abort
5. To exit the Calibration page, press the
continues and calibration progress can be monitored from the
Calibration View).
The Calibrate / Sensor 1 (2, 3, 4) page is displayed.
Calibrate
Sensor 1
Exit
6. Press the
key – the Calibrate page is displayed:
Menu
Calibrate
Exit
7. Press the
key to return to the Operator page.
Yes
ppb
OK
key. (Calibration
Select
Select
CI/AWT540–EN Rev. C

Advertisement

Table of Contents
loading

Table of Contents