Download Print this page

Product Installation - Haier HXC-158B Operation Manual

Blood bank refrigerator
Hide thumbs Also See for HXC-158B:

Advertisement

Specification
Name
Model
External dimensions (W X D X H)
(mm)
Internal dimensions (W X D X H)
(mm)
Effective volume
Door
Insulating materials
Compressor
Shelf / Drawer
5 shelves / 20 baskets
The loading of each shelf
Refrigeration mode
Outer plate / inner plate
Condenser / evaporator
Temperature controller
Refrigerant
Headlamp
Net weight
Temperature of temperature
measurement box
Rated power supply
Rated power / current
Noise Level
Foaming Agent
Climate Type
Electric shock resistance type
Power connect style
USB function / temperature recorder
High-low temperature alarm, sensor fault alarm, power failure
Alarm system
alarm, low battery alarm and door opening alarm
Duration time of power failure alarm
Rechargeable battery
Note: Climate Type 4 means the temperature is +30℃,the relative humidity is 55%.Technical
information may vary from the appliance that you have just purchased due to technical improvements.
38
Blood Bank Refrigerator
HXC-358
HXC-358B
720x690x1730
620x490x1160
358L
Electrical heating glass door
Hard polyurethane foam (fluorine-free)
Totally enclosed imported compressor
5 Drawers
Max:25kg
Forced Wind Cooling Circulation
Spraying steel plate / stainless steel plate
Fin / fin
Computer control system
R134a 420g
LED 3W
158kg
4℃±1℃
220-240V~/50Hz/60Hz
460W/3.0A
43dB(A)
CP/IP
4
I
Y
Standard / optional
48 hours (when battery is fully charged)
DC 12 V rechargeable battery

Product Installation

Installation environment
Ambient temperature: 18°C to 38°C (temperature of HXC-1308/1308B: 16 ~ 32°C), 18°C to 25°C
are optimal and air conditioning system is used as necessary.
Ambient humidity: below 85%Rh.
Avoid excessive dust.
Avoid mechanical swing or vibration.
The refrigerator shall operate at an altitude Lower than 2000m.
Input voltage: rated voltage ± 10%.
As it is sensitive to ambient temperature. Refrigerator can not operate normally
if it is installed in environments other than those described above. Start operation
after the environment is improved.
It is prohibited to install the refrigerator outdoors. Electric leakage or shock may
be caused if the refrigerator gets wet by rainwater.
Installation site
The installation site shall meet following requirements for normal operation and best performance of
refrigerator:
Do not install the refrigerator in a narrow and confined space and the door of the placement room
may not be smaller or lower than that of the refrigerator, to ensure smooth entry and exit of the
refrigerator as well as to prevent difficulty in maintenance and damage to the stored articles due
to not timely repair of refrigerator as the installation site is not easy to be accessed.
The ground at the installation site shall be solid and flat.
It is well ventilated and free from direct sunshine.
Do not share one socket with plugs of other equipment and the plug shall be securely plugged
into the socket.
Do not twist or compress the power line.
If the power line needs to be extended, the extended line shall be no smaller than 3mm² in
sectional area and no longer than 3m.
Check the operating voltage before operation and voltage regulator appropriate for motor load
may be applied in regions with unstable voltage. Ensure the normal input voltage stands at
220V±10% and the power of voltage regulator is greater than 4000W.
The refrigerator shall be grounded in a reliable manner.
If the socket is provided with grounding wire, check if it is well grounded before operation.
If the socket is not provided with grounding wire, a new one shall be installed by engineering
technician.
Do not ground the refrigerator through gas pipes, water supply pipes, telephone
lines or lightning conductors as electric shock may be caused.
The power plug can be accessed after installation to ensure the power line can
be pulled out in a timely manner in emergency cases. The air vent shall be free
from barriers.
7

Hide quick links:

Advertisement

loading