Apply the dedicated power supply marked on the nameplate of refrigerator to prevent fire
or electric shock.
If voltage below 198V or above 242V is applied, automatic voltage regulator above 4000W
and appropriate for motor load shall be installed.
If the power line needs to be extended, the extended line shall be no smaller than 2mm² in
sectional area and no longer than 3m in length. Otherwise, fire or electric shock may be
caused.
The power line for this refrigerator is provided with a standard three wire (grounding) plug,
which complies with 16A. Do not remove or dismantle the grounding pin of power line in
any case. Make sure the power plug is securely and reliably plugged into the socket to
prevent fire.
Apply socket with ground wire to prevent electric shock.
If the socket fails to be grounded, ground wires must be installed by technician.
Do not use the refrigerator outdoors. Electric leakage or shock may be caused if the
refrigerator gets wet by rainwater.
Do not place the refrigerator in humid places or places where the refrigerator may suffer
splashing water. To prevent electric leakage or shock due to deterioration of insulation.
Do not pour water on the refrigerator to prevent electric shock or short circuit.
Do not place containers with water or heavy stuffs on the refrigerator. Personal injuries may
be caused by falling articles and electrical leakage or shock may be caused by out flowed
water due to deterioration of insulation.
Do not ground the refrigerator through gas pipes, water supply pipes, telephone lines or
lightning conductors. As electric shock or other dangers may be caused.
Do not touch electrical parts such as power plug or switches with wet hands to prevent
electric shock.
Hold the power plug rather than the wire when pulling the plug from the socket. As electric
shock or fire due to short circuit may be caused.
Pull out the power plug when the refrigerator is under abnormal performance. As electric
shock or fire may be caused.
Users are not allowed to dismantle, repair or refit the refrigerator by themselves. As fire or
personal injuries may be caused due to improper operation.
Disconnect the refrigerator when repairs or maintenance are performed on the refrigerator
to prevent electric shock or personal injuries.
Make sure medicines or airborne particles inside or around the refrigerator will be not be
inhaled during maintenance as health hazard may be caused.
The refrigerator shall be used in safe regions when toxic, harmful or radioactive articles are
stored inside. As improper use may pose danger to human health or environment.
Pull out the power plug if the refrigerator is out of service for long time to prevent electric
shock, leakage or fire caused by aging power lines.
If the refrigerator is left unused in areas where supervision is unavailable for a long time,
make sure children will not get close to the refrigerator and the door can not be completely
closed.
Do not store inflammable, explosive or volatile articles inside the refrigerator and do not use
flammable sprays nearby as explosion or fire may be caused.
4
Specification
Name
Model
External dimensions (W X D X H)
(mm)
Internal dimensions (W X D X H)
(mm)
Effective volume
Door
Insulating materials
Compressor
Shelf / Drawer
12 shelves / 48 baskets
The loading of each shelf
Refrigeration mode
Outer plate / inner plate
Condenser / evaporator
Temperature controller
Refrigerant
Headlamp
Net weight
Temperature of temperature
measurement box
Rated power supply
Rated power / current
Noise Level
Foaming Agent
Climate Type
Electric shock resistance type
Power connect style
USB function / temperature recorder
High-low temperature alarm, sensor fault alarm, power failure
Alarm system
alarm, low battery alarm and door opening alarm
Duration time of power failure alarm
Rechargeable battery
Note: Climate Type 4 means the temperature is +30℃,the relative humidity is 55%.Technical
information may vary from the appliance that you have just purchased due to technical improvements.
Blood Bank Refrigerator
HXC-1308
HXC-1308B
1440x925x1980
1320x700x1500
1308L
Electrical heating glass door
Hard polyurethane foam (fluorine-free)
Totally enclosed imported compressor
12 Drawers
Max:30kg
Forced Wind Cooling Circulation
Spraying steel plate / stainless steel plate
Fin / fin
Computer control system
R134a 490g
LED 5W
310kg
4℃±1.5℃
220-240V~/50Hz
850W/4.2A
45dB(A)
CP/IP
4
I
Y
Standard / standard
48 hours (when battery is fully charged)
DC 12 V rechargeable battery