General Information - YSI MultiLab IDS 4120 Operation Manual

Ph combination electrodes with refillable liquid reference system
Table of Contents

Advertisement

4120(W) / 4130(W)
Automatic sensor
recognition
Scope of delivery
Commissioning
ba76136e04
10/2017
1

General information

The sensor electronics with the stored sensor data are in the connecting
head of the electrode. The data include, among other things, the sensor type
and series number. In addition, the calibration data are stored in the sensor
with each calibration and the calibration history is recorded (the last 10 cali-
brations). The data is recalled by the meter when the sensor is connected
and is used for measurement and for measured value documentation.
Storing the calibration data in the sensor ensures that the correct slope and
asymmetry are automatically used if the sensor is operated with different
meters. On the other hand, different calibrated sensors can be used with one
meter without the need to recalibrate.
The digital transmission technique guarantees the failure-free communica-
tion with the meter even with long connection cables. The sensor firmware
can be updated via the meter.
2
Commissioning, measuring, calibration
2.1
Commissioning
 Electrode 4120(W) / 4130(W)
 Operating manual
The electrode is filled with reference electrolyte solution in the factory. Pre-
pare the electrode for measuring as follows:
 Open the refilling opening for the reference electrolyte solution. Depend-
ing on the model, the stopper of the refilling opening is an elastomer stop-
per or a slider.
The refilling opening must always be open during calibration and
measurement!
Refilling
opening
 Remove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics
and can easily be removed with deionized water.