YSI MultiLab IDS 4320 Operation Manual

Conductivity cell for ultrapure water

Advertisement

Quick Links

OPERATIONS MANUAL
ba76140e02 10/2017
4320(W)
CONDUCTIVITY CELL FOR ULTRAPURE WATER

Advertisement

Table of Contents
loading

Summary of Contents for YSI MultiLab IDS 4320

  • Page 1 OPERATIONS MANUAL ba76140e02 10/2017 4320(W) CONDUCTIVITY CELL FOR ULTRAPURE WATER...
  • Page 2 4320 Contact 1725 Brannum Lane Yellow Springs, OH 45387 USA Tel: +1 937-767-7241 800-765-4974 Email: environmental@ysi.com Internet: www.ysi.com © 2017 Xylem Inc. Copyright ba76140e02 10/2017...
  • Page 3: Automatic Sensor Recognition

    The digital transmission technique guarantees the failure-free communi- cation with the meter even with long connection cables. If the sensor firmware is enhanced by YSI, it can be updated with the meter. Recommended fields of application Measurements in ultrapure water.
  • Page 4: Measurement Operation

    Measurement / Operation 4320(W) Measurement / Operation Commissioning Scope of delivery  Conductivity measuring cell4320(W)  Operating manual Commissioning Connect the electrode to the meter. 4320 – via the sensor cable to a free IDS connec- tor on the meter 4320W –...
  • Page 5 4320(W) Measurement / Operation Opening and closing the IDS plug connection (4320W) This section only applies to the IDS plug variant, 4320W Plug connection closed and locked Guidance Groove Opening the plug  If necessary, clean the plug connection. connection ...
  • Page 6 Cleaning 4320(W) Cleaning Note To clean the sensor, disconnect it from the instrument. Exterior cleaning We recommend to clean the sensor thoroughly, especially before measuring low conductivity values. Contamination Cleaning procedure Lime sediments Immerse in acetic acid for 5 minutes (volume share = 10 %) Fat/oil Clean with warm water that contains wash-...
  • Page 7 4320(W) What to do if ... What to do if ... Error symptom Cause Remedy No temperature or con- – No connection between meter and – Establish connection between ductivity display conductivity measuring cell meter and conductivity measuring cell – Cable defective Measurement delivers –...
  • Page 8: Technical Data

    Technical data 4320(W) Technical data General data General features Measuring principle 2-electrodes measurement Cell constant 0.100 cm ± 2 % Temperature sensor Integrated NTC 30 (30 k at 25 °C / 77 °F) Dimensions 4320 (in mm) 176,0 120,0 12,0 21,7 4320W 222,8...
  • Page 9 4320(W) Technical data Connection cable Lengths 4320: 1,5 m 4320W: 1,5 / 3 m Diameter 4.3 mm Smallest allowed Fixed installation: 20 mm bend radius Flexible use: 60 mm Plug type Socket, 4 pins IDS plug Type of connection 4-pole, watertight plug connection with lock, reverse polarity protected (4320W) Materials...
  • Page 10 Technical data 4320(W) Characteristics Temperature responding (99 % of the final value display after) when delivered behavior < 20 s Accuracy of the temperature ± 0.2 K sensor Measuring ranges and resolution Measuring ranges, Measured parameter Measuring range Resolution resolution ӕ...
  • Page 11: Wear Parts And Accessories

    4320(W) Wear parts and accessories Wear parts and accessories Connection cable Description Model Order no. 4320W IDS connection cable, 1.5 m IDS-CABLE- 903 850Y - meter IDS connection cable, 3 m IDS-CABLE-3 903 851Y Radio connection Description Model Order no. 4320W - meter WA capable IDS meter see Internet...
  • Page 12 Wear parts and accessories 4320(W) ba76140e02 10/2017...
  • Page 13: Contact Information

    When returning a product for service, include the Product Return form with cleaning certification. The form must be completely filled out for an YSI Service Center to accept the instrument for service. The Product Return form may be downloaded at www.ysi.com...
  • Page 14 Contact Information 4320 ba76140e02 10/2017...
  • Page 16 For more information on how Xylem can help you, go to www.xyleminc.com 1725 Brannum Lane Yellow Springs, OH 45387 Tel: +1 937-767-7241; 800-765-4974 Fax: +1 937-767-1058 Email: environmental@ysi.com Web: www.ysi.com ©Xylem Inc...

This manual is also suitable for:

Multilab ids 4320w