YSI MultiLab IDS 4310 Operation Manual

Conductivity cell

Advertisement

Quick Links

OPERATIONS MANUAL
ba76139e02 10/2017
4310(W)
CONDUCTIVITY CELL

Advertisement

Table of Contents
loading

Summary of Contents for YSI MultiLab IDS 4310

  • Page 1 OPERATIONS MANUAL ba76139e02 10/2017 4310(W) CONDUCTIVITY CELL...
  • Page 2 4310(W) Contact 1725 Brannum Lane Yellow Springs, OH 45387 USA Tel: +1 937-767-7241 800-765-4974 Email: environmental@ysi.com Internet: www.ysi.com © 2017 Xylem Inc. Copyright ba76139e02 10/2017...
  • Page 3: Automatic Sensor Recognition

    The digital transmission technique guarantees the failure-free communication with the meter even with long connection cables. If the sensor firmware is enhanced by YSI, it can be updated via the meter. Recommended fields of application  On site measurements in rivers, lakes and wastewater ...
  • Page 4: Measuring / Operation

    Measuring / Operation 4310(W) Due to the use of high performance PEEK material for the connection technology 4310 has an improved chemical resistance, especially in acid media (eg. As electroplating baths). NOTE Concentrated or strong oxidizing acids and organic solvents can damage or destroy the sensor.
  • Page 5 4310(W) Cleaning Plug connection closed and locked Guidance Groove Opening the plug  If necessary, clean the plug connection connection  Open the locking device (step 1)  Use your thumb and index finger to press the clips of the connector together, and pull the connector out of the plug (step 2).
  • Page 6 What to do if... 4310(W) Contamination Cleaning procedure Lime sediments Immerse in acetic acid for 5 minutes (volume share = 10 %) Fat/oil Clean with warm water that contains wash- ing-up liquid After cleaning, thoroughly rinse with deionized water and recalibrate if necessary.
  • Page 7: Technical Data

    4310(W) Technical data Technical data General data General features Measuring principle Four-electrodes measurement Cell constant 0.475 cm ±1.5 % Temperature sensor Integrated NTC 30 (30 k at 25 °C / 77 °F) Dimensions (in mm) 4310 176,0 120,0 15,3 21,7 4310W 222,8 120,0...
  • Page 8 Technical data 4310(W) Smallest allowed Fixed installation:20 mm bend radius Flexible use:60 mm Plug type Socket, 4 pins Connection type 4-pole, watertight plug connection with lock, IDS plug (4310W) reverse polarity protected Materials  Synthetic materials: Glass fiber reinforced Noryl, TPU, TPC-ET, POM, PEEK, PBT ...
  • Page 9: Measuring Ranges, Resolution, Accuracy

    4310(W) Technical data Measuring ranges, resolution, accuracy Measuring ranges, Measured parameter Measuring range Resolution resolution ӕ [µS/cm] 0.0 ... 199.9 200 ... 1999 ӕ [mS/cm] 2.00 ... 19.99 0.01 20.0 ... 199.9 200 ... 2000 ρ ( Resistivity) [Ohm*cm] 0.50 ... 19.99 0.01 20.0 ...
  • Page 10: Wear Parts And Accessories

    Wear parts and accessories 4310(W) Wear parts and accessories Connection cable Description Model Order no. 4310W IDS connection cable, 1.5 m IDS-CABLE- 903 850Y - meter IDS connection cable, 3 m IDS-CABLE-3 903 851Y Radio connection Description Model Order no. 4310W - meter WA capable IDS meter see Internet...
  • Page 11: Contact Information

    When returning a product for service, include the Product Return form with cleaning certification. The form must be completely filled out for an YSI Service Center to accept the instrument for service. The Product Return form may be downloaded at www.ysi.com...
  • Page 12 Contact Information 4310 ba76139e02 10/2017...
  • Page 14 For more information on how Xylem can help you, go to www.xyleminc.com 1725 Brannum Lane Yellow Springs, OH 45387 Tel: +1 937-767-7241; 800-765-4974 Fax: +1 937-767-1058 Email: environmental@ysi.com Web: www.ysi.com ©Xylem Inc...