Commissioning, measuring, checking Commissioning Scope of delivery Electrode 4210W or 4211 Operating manual Commissioning Prepare the electrode for measuring as follows: Open the refilling opening for the reference electrolyte solution. Depend- ing on the model, the stopper of the refilling opening is an elastomer stop- per or a slider.
Page 4
Commissioning, measuring, checking 4210W / 4211 4210W – via a connecting cable (accessory) to a free IDS connector on the meter – wireless via an IDS WA-S adapter (acces- sory) to a WA-capable meter Accessories for the connection of the 4211W...
4210W / 4211 Commissioning, measuring, checking Opening and closing the IDS plug connection This section only applies to variants with IDS plug (4210W). Plug connection closed and locked Guidance Groove Opening the plug If necessary, clean the plug connection connection ...
Commissioning, measuring, checking 4210W / 4211 Measuring: General rules Make sure the refilling opening for the reference electrolyte solution is open. Avoid the carryover of sample solution from one measurement to the next as follows: – Shortly rinse the sample beakers with the solution the beakers are to be filled with next.
Page 7
4210W / 4211 Storage T (°C) [mV] T (°C) [mV] 4210W / 4211 4210W / 4211 +214 +188 +211 +184 +207 +180 +203 Checking with URef is temperature dependent and can be taken from the following table: Zobell ORP buffer solution T (°C)
Junction Ceramic Double junction Electrode material and shape Platinum / circle Connection cable Lengths 4211: 1.5 m 4210W: 1.5 / 3 m Diameter 4.3 mm Smallest allowed Fixed installation:20 mm bend radius Flexible use:60 mm Plug type Socket, 4 -Ppoles...
Wear parts and accessories 4210W / 4211 IDS plug Connection type 4-P-pin, watertight plug connection with locking device, reverse polarity protected Storage With watering cap; filled with KCl 3 mol/L, Ag free Wear parts and accessories Maintenance Description Model Order no.
When returning a product for service, include the Product Return form with cleaning certification. The form must be completely filled out for an YSI Service Center to accept the instrument for service. The Product Return form may be downloaded at www.ysi.com...