YSI 4210W Operation Manual

Orp electrode

Advertisement

Quick Links

OPERATIONS MANUAL
ba76138e05 06/2018
4210W
4211
ORP ELECTRODE

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 4210W and is the answer not in the manual?

Questions and answers

Summary of Contents for YSI 4210W

  • Page 1 OPERATIONS MANUAL ba76138e05 06/2018 4210W 4211 ORP ELECTRODE...
  • Page 2 4210W / 4211 For the most recent version of the manual, please visit www.ysi.com. Contact 1725 Brannum Lane Yellow Springs, OH 45387 USA Tel: +1 937-767-7241 800-765-4974 Email: info@ysi.com Internet: www.ysi.com © 2018 Xylem Inc. Copyright ba76138e05 06/2018...
  • Page 3: Automatic Sensor Recognition

    Commissioning, measuring, checking Commissioning Scope of delivery  Electrode 4210W or 4211  Operating manual Commissioning Prepare the electrode for measuring as follows:  Open the refilling opening for the reference electrolyte solution. Depend- ing on the model, the stopper of the refilling opening is an elastomer stop- per or a slider.
  • Page 4 Commissioning, measuring, checking 4210W / 4211 4210W – via a connecting cable (accessory) to a free IDS connector on the meter – wireless via an IDS WA-S adapter (acces- sory) to a WA-capable meter Accessories for the connection of the 4211W...
  • Page 5: Opening And Closing The Ids Plug Connection

    4210W / 4211 Commissioning, measuring, checking Opening and closing the IDS plug connection This section only applies to variants with IDS plug (4210W). Plug connection closed and locked Guidance Groove Opening the plug  If necessary, clean the plug connection connection ...
  • Page 6: Measuring: General Rules

    Commissioning, measuring, checking 4210W / 4211 Measuring: General rules  Make sure the refilling opening for the reference electrolyte solution is open.  Avoid the carryover of sample solution from one measurement to the next as follows: – Shortly rinse the sample beakers with the solution the beakers are to be filled with next.
  • Page 7 4210W / 4211 Storage T (°C) [mV] T (°C) [mV] 4210W / 4211 4210W / 4211 +214 +188 +211 +184 +207 +180 +203 Checking with URef is temperature dependent and can be taken from the following table: Zobell ORP buffer solution T (°C)
  • Page 8: Maintenance And Cleaning

    Aging 4210W / 4211 Aging ORP electrodes are consumables. Every ORP electrode undergoes a natural aging process. Extreme operating conditions can considerably shorten the lifetime of the electrode. These are:  Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bro- mides, sulfides, iodides, proteins ...
  • Page 9: Technical Data

    Junction Ceramic Double junction Electrode material and shape Platinum / circle Connection cable Lengths 4211: 1.5 m 4210W: 1.5 / 3 m Diameter 4.3 mm Smallest allowed Fixed installation:20 mm bend radius Flexible use:60 mm Plug type Socket, 4 -Ppoles...
  • Page 10: Wear Parts And Accessories

    Wear parts and accessories 4210W / 4211 IDS plug Connection type 4-P-pin, watertight plug connection with locking device, reverse polarity protected Storage With watering cap; filled with KCl 3 mol/L, Ag free Wear parts and accessories Maintenance Description Model Order no.
  • Page 11: Contact Information

    When returning a product for service, include the Product Return form with cleaning certification. The form must be completely filled out for an YSI Service Center to accept the instrument for service. The Product Return form may be downloaded at www.ysi.com...
  • Page 12 For more information on how Xylem can help you, go to www.xyleminc.com 1725 Brannum Lane Yellow Springs, OH 45387 Tel: +1 937-767-7241; 800-765-4974 Fax: +1 937-767-1058 Email: info@ysi.com Web: www.ysi.com ©Xylem Inc...