Advertisement

Technical data
Characteristics
when delivered
Measuring ranges,
resolution
10
Temperature responding
behavior
Accuracy of the temperature
sensor
5.2
Measuring ranges and resolution
Measured parameter
ӕ [µS/cm]
ρ (resistivity) [kOhm*cm]
ρ (resistivity) [MOhm*cm]
T [°C]
5.3
Accuracy of the IDS measuring technique
Measured parameter
ӕ, ρ
T [°C]
t
(99 % of the final value display after)
99
< 20 s
± 0.2 K
Measuring range
0.01 ... 19.99
0.0 ... 199.9
5,00 ... 19,99
20.0 ... 199.9
200 ... 1999
2.00 ... 19.99
20.0 ... 199.9
− 5,0 ... + 100,0
Accuracy (± 1 digit)
± 0.5 % of measured value
± 0.1
4320(W)
Resolution
0.01
0.1
0.01
0.1
1
0.01
0.1
0.1
ba76140e02