  • Page 1 HELiOS Reservoirs User Manual (English)
  • Page 2: Symbols Glossary

    HELiOS Symbols Glossary This device complies with the re- ISO 7000; Graphical symbols for use on equip- quirements of Directive 2010/35/EU ment—Index and synopsis concerning medical devices. It bears Operating temperature limitation of the pi marking as shown. these units is 10°C to 40°C. Storage ADR: European Agreement concerning the Inter- temperature limitation range is -40°C national Carriage of Dangerous Goods by Road...
  • Page 3: Specifications

    12.5 mm in diameter and ingress of vertically dripping water. • Equipment not suitable for use in the presence of flammable mixtures Product Specifications HELiOS U36 HELiOS U46 HELiOS 36 HELiOS 46 LOX capacity...
  • Page 4: Warning Information

    HELiOS Warning Information Liquid and gaseous oxygen, though nonflammable, cause other materials to Important: Read this manual thoroughly before burn faster than normal. This hazard, operating the HELiOS. along with the low temperature of liquid RX Only. oxygen, warrants certain safety precau- tions.
  • Page 5 The HELiOS reservoir stores liquid oxygen that can be used to fill a portable liquid oxygen tank. HELiOS Universal reservoirs can fill any CAIRE top fill (TF) portable with a compatible fill connec- PN 21157307 A | User Manual...
  • Page 6 HELiOS H300 and H850 portable units. • If the yellow low battery light appears, inform your healthcare provider the next time your HELiOS Universal reservoirs will fill any CAIRE HELiOS reservoir is filled. top-fill (TF) portable unit that has a compatible NOTE: The HELiOS reservoir will continue to fill and female fill connector.
  • Page 7: Maximum (Recommended)

    HELiOS 2. Use the following chart as a guideline to the Operating Instructions – recommended tubing length. Filling a Portable FLOW MAXIMUM (RECOMMENDED) SETTING TUBING LENGTH* (LPM) 22-psig 1. Verify the level of liquid oxygen in the unit (see Operating Instructions — Level Verification). 100 Ft.
  • Page 8 HELiOS 5. Turn the knob on the external flow control valve Operating Instructions – until your prescribed continuous flow rate (nu- Using the HELiOS Supply Line meric in LPM) is displayed in the window and a positive detent is felt. NOTE: The knob should not be set higher than the 1.
  • Page 9: Maintenance

    HELiOS Maintenance Clean the fill connectors on both the stationary Your service provider is responsible for any mainte- and portable units with a clean, dry, lint-free cloth nance that my be required per the technical manual between each fill to prevent freezing and possible of this device.
  • Page 10: Weee And Rohs

    HELiOS Cleaning Standard Accessories WARNING: CLEAN ONLY AFTER THE UNIT 15 in (38.1 cm) Roller Base for IS EMPTY. HELiOS Reservoirs, available for all models. • Clean using a solution of mild dish washing detergent and water. NOTE: Only use roller base on flat •...
  • Page 11: Guidance And Manufacturer's Declaration - Electromagnetic Emissions

    HELiOS Safety WARNING: PORTABLE RF COMMUNICATIONS EQUIPMENT (INCLUDING PERIPHERALS SUCH AS ANTENNA CABLES AND EXTERNAL ANTENNAS) SHOULD BE USED NO CLOSER THAN 30 CM (12 INCHES) TO ANY PART OF THE LIBERATOR, INCLUDING CABLES SPECI- FIED BY THE MANUFACTURER. OTHERWISE, DEGRADATION OF THE PERFORMANCE OF THIS EQUIPMENT COULD RESULT.
  • Page 12: Guidance And Manufacturer's Declaration - Electromagnetic Immunity

    HELiOS Table 2* Recommended separation distances between portable and mobile RF communications equipment and the HELiOS Reservoir The HELiOS Reservoir is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the HELiOS Reservoir can help prevent elec- tromagnetic interference by maintaining a minimum distance between portable and mobile RF commu- nications equipment (transmitters) and the HELiOS Reservoir as recommended below, according to the maximum output power of the communications equipment.
  • Page 13 HELiOS Table 4 Guidance and Manufacturer’s Declaration—Immunity ME Equipment and ME Systems Guidance and Manufacturer’s Declaration—Immunity The HELiOS Reservoir is intended for use in the electromagnetic environment specified below. The customer or user of the HELiOS Reservoir should ensure that it is used in such an environment. Immunity Test IEC 60601 Test Level Compliance Level...
  • Page 14 HELiOS Notes 14 - ENG PN 21157307 A | User Manual...
  • Page 15 HELiOS Notes PN 21157307 A | User Manual 15 - ENG...
  • Page 16 Tel/Fax: (55)(11)5535-0989 email: medstar@medstar.com.br Copyright © 2018 CAIRE Inc. CAIRE Inc. reserves the right to discontinue its products, |21157307_Aa~ or change the prices, materials, equipment, quality, descriptions, specifications and/or, processes to its products at any time without prior notice and with no further obligation or 21157307 Rev A consequence.