Siemens 7MC3052-1AB68 Operating Instructions Manual page 53

Two-colour pyrometer, 7mc3052-series
Table of Contents

Advertisement

7MC3052-1AE66
Temperature range
Sensor
Spectral sensitivity
Focus distance
Target spot diameter
Analogue output
Resolution current output
Resolution temp. reading
Response time t
90
Repeatability
Measurement uncertainty
Temperature coefficient
Ambient temperature
Current consumption
Dimensions
Energy
95 %
0.66 0.68
PZ41.18
(33 - 45 mm)
33
700 - 1800 °C
Double Si
0.95 / 1.05 µm
33 - 45 mm
50 : 1
0(4) - 20 mA linear, switchable, scalable (≥ 50 K)
0.2 K + 0.03 % of selected range
1 K
≤ 10 ms (for T > 800 °C)
2 K
1.0 % of temp. reading [°C] plus 3.0 K
0.05 %/K of temp. reading / K
(deviation to Tu = 23 °C)
0 - 65 °C
≤ 50 mA at 24 V DC without load current
M30 x 200 mm (without plug)
Target diameter [mm]
0.7
0.72 0.74 0.76 0.78
34
35
36
37
Target distance [mm]
0.8
0.82 0.84 0.86 0.88
38
39
40
41
47
Operating instructions
0.9
45
42
43
44

Advertisement

Table of Contents
loading