Appendix C. Updating the Firmware in the Field
C.2 Performing the Firmware Update
If your AT600 flow meter system meets all of the requirements discussed on the previous page and you are
prepared to perform the firmware update according to those guidelines, proceed with the instructions in this
section.
C.2.1 Check the Current Firmware Version
To determine the firmware version currently installed in your AT600, access the following information
screen:
Main Menu > Program > Advanced > Flow Meter Data > Main Board > SW Version
For reference, an example of this screen is shown in Figure 36 below.
Note: As indicated on the previous page, the current version of your AT600 firmware must be 01.02.23 or later.
If your version is older than this, you cannot use this update method. Software versions 01.02.24,
01.03.xx, 02.xx.xx are examples of acceptable versions.
C.2.2 Update Steps
If your AT600 firmware version qualifies for the field firmware update process, complete the following steps:
1. Prepare the RS485 Modbus connection:
a. Disconnect the main power from the AT600.
b. Wire the Modbus connection as described in "Wiring the Modbus Communication" on page 28.
2. Locate the AquaTrans Flow Meter Software Update (version 20161117V1.2 or later) software on your
PC. If the folder containing the software is zipped, you must unzip it prior to use.
184
SW Version
01.02.23
Figure 36: Example Software Version Screen
AquaTrans™ AT600 User's Manual