ResMed S9 Elite Information Manual page 5

Hide thumbs Also See for S9 Elite:
Table of Contents

Advertisement

Available languages

Available languages

Electromagnetic compatibility
Aircraft use
IEC 60601-1 classification
S9 technical specifications
Operating pressure range
Maximum single fault steady state pressure
Pressure measurement tolerance
Flow measurement tolerance
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:
Sound pressure level
Sound power level
Nominal dimensions (L x W x H)
Weight
Housing construction
Supplemental oxygen
Hypoallergenic air filter
Standard air filter
Air outlet
H5i technical specifications
Maximum heater plate temperature
Temperature cut-out
Maximum gas temperature
Nominal dimensions (L x W x H)
Weight (water tub)
Weight (cleanable water tub)
Water capacity
Docking station
Product complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential,
commercial, and light industry environments. Information regarding the
electromagnetic emissions and immunity of these ResMed devices can
be found on www.resmed.com, on the Products page under Service
and Support. Click on the PDF file for your language.
ResMed confirms that the S9 AutoSet/Elite meets the Federal Aviation
Administration (FAA) requirements (RTCA/DO-160F) for all phases of air
travel.
Class II (double insulation), Type BF
4 to 20 cm H
O
2
30 cm H
O
2
±0.5 cm H
O ± 4% of the measured reading
2
±6 L/min or 10% of reading, whichever is greater
24 dBA as measured according to ISO 17510-1:2002
26 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1:2007
34 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1:2007
6.0" x 5.5" x 3.4" (153 mm x 140 mm x 86 mm)
1.8 lb (810 g)
Flame retardant engineering thermoplastic
Recommended maximum supplemental oxygen flow: 4 L/min
Non-woven acrylic and polypropylene fibers with polypropylene carrier
Polyester fiber
The 22 mm conical air outlet complies with ISO 5356-1
150°F (65°C)
165°F (74°C)
≤ 106°F (≤ 41°C)
Docking station and water tub: 6.0" x 5.7" x 3.4"
(153 mm x 145 mm x 86 mm)
Docking station and unfilled water tub 1.5 lb (0.67 kg)
Docking station and unfilled water tub 1.7 lb (0.77 kg)
To maximum fill line 380 mL
Flame retardant engineering thermoplastic, aluminium
5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

S9 autosetH5i

Table of Contents