Problem/Possible Cause
SmartStart/Stop is enabled but the CPAP device does not stop automatically when you remove your mask
Incompatible mask system being used.
The patient is using a nasal pillows mask with a
set pressure less than 7 cm H
Pressure rises inappropriately
Talking, coughing or breathing in an unusual
manner.
Mask cushion is buzzing against the skin.
Displays error message: Heated tube fault, replace tube
Device has been left in a hot environment.
There is a fault in your ClimateLine air tubing.
Displays error message: Humidifier fault, replace humidifier
Device has been left in a hot environment.
There is a fault in your H5i heated humidifier.
Displays error message: High temperature fault, refer to user manual
Device has been left in a hot environment.
The device or air inlet is being covered by
bedding.
Air filter is blocked.
Air tubing is blocked.
Humidifier setting is too high, resulting in
accumulation of water in the air tubing.
Displays error message: Check ResMed Power Supply Unit and fully insert the connector
The DC plug is partially inserted into the back of
the device or inserted too slowly.
A non-ResMed power supply unit is connected to
the device.
The power supply unit is being covered by
bedding.
Solution
Only use equipment recommended by ResMed.
Disable SmartStart/Stop.
O.
2
Avoid talking with a nasal mask on, and breathe as normally as possible.
Adjust the headgear.
Allow to cool before re-use. Disconnect the power cord and then
reconnect it to restart the device.
Discontinue using your ClimateLine air tubing and contact your clinician/
service provider. Use Standard or SlimLine tubing in the interim.
Allow to cool before re-use. Disconnect the power cord and then
reconnect it to restart the device.
Discontinue using your H5i and contact your clinician/service provider.
Allow to cool before re-use. Disconnect the power cord and then
reconnect it to restart the device.
Make sure that the device is free from bedding, clothes or other objects
that could cover the device or block the air inlet. Disconnect the power
cord and then reconnect it to restart the device.
Replace your air filter. Disconnect the power cord and then reconnect it to
restart the device.
Check your air tubing and remove any blockages. Disconnect the power
cord and then reconnect it to restart the device.
Turn the humidifier setting down and empty the water from the air tubing.
Fully insert the DC plug.
Remove the power supply unit and replace with a ResMed power supply
unit.
Make sure that the power supply unit is free from bedding, clothes or
other objects that could cover it.
3