Watermaster Flow Performance; Watermaster Flow Performance - M3/H - ABB WaterMaster Manual

Electromagnetic flowmeter
Hide thumbs Also See for WaterMaster:
Table of Contents

Advertisement

WaterMaster
Electromagnetic flowmeter

6 WaterMaster Flow Performance

6.1 WaterMaster flow performance – m
Q4
3
DN
(m
/h)
10
3.1
15
7.88
20
12.5
25
20
32
31.25
40*
50
50*
79
65*
125
80*
200
100*
313
125*
313
150*
788
200*
1,250
250
2,000
300
3,125
350
5,000
400
5,000
450
7,875
500
7,875
600
12,500
10,000
700
20,000
16,000
30 in
20,000
16,000
800
20,000
16,000
900
31,250
25,000
36
Standard Calibration
0.4 % Class 2
Q3
Q
0.4%
3
3
(m
/h)
(m
/h)
2.5
0.167
6.3
0.42
10
0.67
16
1.1
25
1.67
40
4.2
63
4.2
100
6.7
160
10.7
250
16.7
250
16.7
630
42
1,000
67
1,600
107
2,500
167
4,000
267
4,000
267
6,300
420
6,300
420
667
1600
1600
1600
2500
3
/h
Q2
Q1
Q
3
3
(m
/h)
(m
/h)
(m
0.013
0.008
0.032
0.02
0.05
0.032
0.08
0.05
0.13
0.08
0.2
0.13
0.32
0.20
0.5
0.32
0.81
0.51
1.3
0.79
1.3
0.79
3.2
2.0
5.1
3.2
8.1
5.1
12.7
7.9
20.3
12.7
20.3
12.7
32
20
32
20
51
32
1000
102
64
1600
102
64
1600
102
64
1600
160
100
2500
6 WaterMaster Flow Performance
High Accuracy Calibration
0.2 % Class 1
Q2
0.2%
3
3
/h)
(m
/h)
0.31
0.02
0.79
0.05
1.25
0.08
2
0.13
3
0.20
6
0.32
7.9
0.5
12.5
0.8
16
1.3
25
2
25
2
63
5
100
8
160
13
250
20
400
32
400
32
630
50
630
50
80
160
160
160
250
IM/WM–EN Rev. M
Q1
3
(m
/h)
0.012
0.03
0.05
0.08
0.13
0.2
0.32
0.5
0.8
1.25
1.25
3.2
5
8
12.5
20
20
32
32
50
100
100
100
156

Hide quick links:

Advertisement

Table of Contents
loading