  • Page 1 OCC-CWDM-301-18 Optical Channel Checker ERIES Operation Guide...
  • Page 2: Safety Instructions

    OCC-CWDM-301-18 Notices No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into another language) without prior agreement and written consent of the Provider, as governed by international copyright laws. Warranty The material contained in this document is subject to change without notice.
  • Page 3 OCC-CWDM-301-18 Safety Terms Used in This Manual The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personnel injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.
  • Page 4: Table Of Contents

    OCC-CWDM-301-18 Table of Contents 1. General Information ............................6 1.1 Scope of this Manual ............................6 1.2 Unpacking and Inspection ..........................6 1.3 Introduction ..............................6 2. Basic Operation ..............................8 2.1 Foreword ................................8 2.2 Instrument Interfaces Instructions ........................8 2.3 Use of Batteries ..............................9 2.4 Keypad Functions ............................10 3.
  • Page 5 OCC-CWDM-301-18 4.3.1 Switch between Information Window Items ...................37 4.3.2 Review Event List ...........................37 4.3.3 Review Marker A/B Information ......................37 4.4 Zoom in Trace Horizontally ...........................37 4.5 Zoom out Trace Horizontally .........................38 4.6 Zoom in Trace Vertically ..........................38 4.7 Zoom out Trace Vertically ..........................38 4.8 Re-analyze the trace ............................38 4.9 Save Trace ..............................39 4.10 Browse Saved Traces ...........................40...
  • Page 6: General Information

    301-18 Series is the ideal tool for CWDM installation and maintenance technicians. Features: Basic applications: • Clear TFT LCD display (320×240) • 18-channel measurement: Model OCC-CWDM-301-18B, 1271-1611nm • 8-channel measurement: Model OCC-CWDM-301-18A, 1471-1611nm • Result display in histogram and list •...
  • Page 7 OCC-CWDM-301-18 • Backlight • 10 hours continuous operation • Pocketsize, lightweight and easy-to-use • CE, FCC certificates Result display in histogram and list Straightforward result display for easy understanding. Internal clock & fiber S/N editable Internal clock enables OCC-50 to save test data with time and editable fiber SN information for convenient archiving and editing.
  • Page 8: Basic Operation

    OCC-CWDM-301-18 2. Basic Operation 2.1 Foreword This part introduces the basic operation on the OTDR. Specific operations of each type instrument are elaborated in Chapter 3 of this manual. Please read this manual carefully for optimal operation. Should you encounter any problems during operation, you are welcome to contact the technical staff of our company or representatives.
  • Page 9: Use Of Batteries

    OCC-CWDM-301-18 2.3 Use of Batteries Battery for OTDR is NiMH battery. Cautions during Operation: The following may bring auto power off of the instrument: • The instrument will be auto power off when there is insufficient power during operation and low power will be shown on the LCD.
  • Page 10: Keypad Functions

    OCC-CWDM-301-18 2.4 Keypad Functions Power on or off [On/Off] Under GUI, Press to start measurement. While testing, press this key to stop measurement Under GUI, press this key to confirm the current operation Enter Together with [Shift/ ], can browse the event list downwards Move menu bar in menu operation Highlight the icon to be operated ▲...
  • Page 11: Basic Information Of Otdr

    OCC-CWDM-301-18 3. Basic Information of OTDR 3.1 Principle of OTDR OTDR (Optical Time Domain Reflectometer) is a measurement instrument for identifying optic fiber transmission features. The instrument is mainly used to measure attenuation of a whole optic fiber chain and provide attenuation details relating to length, namely detect, locate and measure any event in optic fiber chain (events refer to faults caused by welding, connectors, and bending whose transmission change can be measured).
  • Page 12 OCC-CWDM-301-18 3.2.1.1 Reflection Events When some pulse energy is scattered, reflection events happen. When reflection event occurs, peak shows on trace, as shown in Figure 3-1. Figure 3-1. Reflection Event 3.2.1.2 Non-Reflection Events Non-reflection events happen at certain points where there is some optic loss but no light scattering. When non- reflection event occurs, a power decline shows on trace, as in Figure 3-2.
  • Page 13: Measurement Application Of Otdr

    OCC-CWDM-301-18 3.3 Measurement Application of OTDR OTDR displays power relating to distance of returning signals. This information can be used to identify the main properties of an optic fiber chain. 3.3.1 Measurement Contents of OTDR • Event location (distance), end or rupture of optic fiber chain •...
  • Page 14: Trace Display Of Otdr

    OCC-CWDM-301-18 3.4.1 Trace Display of OTDR This window displays the trace after one measurement. Definition of Trace: After one measurement, reflection power diagram will be displayed as distance function. This diagram is referred to as trace. Trace of OTDR displays measurement result in a graphic form. The y-axis stands for power, and the x-axis stands for distance, as shown in Figure 3-4.
  • Page 15 OCC-CWDM-301-18 8K m /D i v 5.00 dB /D i v t ra c e1 P a ra -2 D a t e : 27- M a y- 2005 09 :15 E nd T hre .: N Re fl .T hre .: 0 .20 3.00 Re fl .T hre .:...
  • Page 16 OCC-CWDM-301-18 Cuml: Cumulative loss, calculating from beginning point to the current event. 3.4.2.3 Information of Marker A/B Marker is used to mark and analyze a single event, trace section and distance. Distance, attenuation, loss at marker or between markers will be displayed in information of markers, as shown in Figure 3-7. 8 K m/ D i v 5.00 dB/ D i v t ra c e1 Ma rke r...
  • Page 17: Menu Bar And Window Of Otdr

    OCC-CWDM-301-18 Figure 3-8. Information of Fiber 3.4.3 Menu Bar and Window of OTDR 3.4.3.1 Menu Bar and Icons of OTDR Icon Description Parameter configuration Save file Open file Re-analyze the trace Zoom in trace horizontally Zoom out trace horizontally Zoom in trace vertically Zoom out trace vertically Switching between markers Review events list upwards...
  • Page 18: Occ-Cwdm

    OCC-CWDM-301-18 3.4.3.2 Parameter Configuration on OTDR Menu Bar Correct parameter configuration is a necessity for accurate measurement; therefore, necessary configuration must be performed before using the instrument. Please refer to the following examples: Use [▲] and [▼] to highlight, i.e. parameter configuration, then press [Enter], as shown in Figure 3-9; Press [ ] to exit.
  • Page 19 OCC-CWDM-301-18 Definitions of Measurement Parameters Parameter Definition of Parameter Length of optic fiber relevant to the trace Range Width of laser pulse sending out from OTDR to optic fiber Pulse Width Average Time To select suitable testing time. To select laser wave length for measurement Wave length Measurement Mode To select mode for measurement...
  • Page 20 OCC-CWDM-301-18 Range Configuration Generally, range will be set according to actual length of optic fiber, so as to insure the accuracy of measurement. Under the menu of parameter configuration, use [▲] and [▼] to highlight “Range”; Press [Enter] to enter, as shown in Figure 3-10;...
  • Page 21 OCC-CWDM-301-18 R a n g e A ut o 1 5 s A ut o P u l s e W i d t h 30 s A v g . T i m e 1 m i n W a v e l e n g t h 1550 nm 2 m i n M e a s .
  • Page 22 OCC-CWDM-301-18 Average Time Configuration Average time will affect the SNR directly. The longer the average time is, the higher SNR is, as well as dynamic range. Therefore, in case of measurement of long-distance optic fiber, long average time should be selected in order to review events at long-distance end.
  • Page 23 OCC-CWDM-301-18 Measurement Mode Configuration There are two kinds of measurement mode: Averaging and Real time mode. Under Real time Mode, OTDR will undertake realtime measurement for the connector of exterior fiber and refurbish the measure trace. While under Real time Mode, press key “ ”...
  • Page 24 OCC-CWDM-301-18 Length Units Configuration Under the parameter configuration menu use [▲] and [▼] to highlight “Length Units”; press [Enter] to select the desired units of measurement, as in Figure 3-16. Press [ ] to exit R a n g e A ut o A ut o P u l s e W i d t h...
  • Page 25 OCC-CWDM-301-18 Scatter Coefficient Configuration Scatter coefficient determines the value of backward scatter power. The configuration affects the calculation of reflection value. Under parameter configuration, use [▲] and [▼] to highlight “Scatter coefficient”; press [Enter] to enter, as shown in Figure 3-18. Press [ ] to exit.
  • Page 26 OCC-CWDM-301-18 Use [◄] and [►] to adjust the position of highlights; use [▲] and [▼] to change the digits. After setting, press [Enter] to confirm. Note The default setting is 0.2dB Reflection Threshold Configuration This configuration has direct impact on reflection events listing. Only reflection events GE this threshold will be displayed in events list.
  • Page 27 OCC-CWDM-301-18 End Threshold Configuration This threshold is the end threshold of optic fiber. If the end threshold equals 3.0dB, then the first event with insertion loss GE 3dB will be considered as the end of the optic fiber. If the value is set to 0dB, there will be no end threshold.
  • Page 28 OCC-CWDM-301-18 Press [▲] and [▼] to choose the files to be deleted, and then press [Enter] to confirm. Users can delete one or several files by one time. Press [◄] and [►] to choose [Delete]. Press [Enter], according to the instruction, choose “Yes”...
  • Page 29 OCC-CWDM-301-18 Note The default setting is “auto off” on Language Configuration There are two language options: English and Chinese. Under parameter configuration, use [▲] and [▼] to highlight “Lang./语言”; press [Enter] to switch, as shown in Figure 3-25. Press [ ] to exit.
  • Page 30 OCC-CWDM-301-18 Color Mode Setting Choose the different displaying color scheme according to the user’s fancy. Under parameter configuration, use [▲] and [▼] to highlight “Color mode”, press [Enter] to choose different mode. Press [ ] to exit, as shown in Figure 3-27. L e n g t h U n i t s M e t e r [ m ] I O R 1 .4666...
  • Page 31 OCC-CWDM-301-18 Help Users can get the quick reference via Help menu. Under parameter configuration, Use [▲] and [▼] to highlight “Help”; Press [Enter] to enter, as shown in 3-29.(a), (b), (c). Press [ ] to exit. S c a t . C o e f . -52.
  • Page 32: Battery Recharge Status

    OCC-CWDM-301-18 3.5 Battery Recharge Status When the instrument is power off and powered through AC/DC adapter, the “CHARGE” indicator on the coping (Figure 1.) will turn on. When batteries are fully recharged, the indicator will turn off. When the instrument is power on and powered through AC/DC adapter, the inside batteries are automatically recharged.
  • Page 33: Trace Measurement And Processing Of Existing Traces

    OCC-CWDM-301-18 4. Trace Measurement and Processing of Existing Traces 4.1 Instructions on GUI After power on, power on interface displays on the LCD, as shown in Figure 4-1: Figure 4-1. Power on Interface OTDR-XXXX: Model ---Version X.X---: Software Version. Three seconds after power on, the interface will be automatically directed to quick reference: 1 .
  • Page 34: Trace Measurement Of Otdr

    OCC-CWDM-301-18 4.2 Trace Measurement of OTDR One complete trace can be obtained for each measurement. Also, OTDR can load a saved trace. Note • Before each measurement, if the operator is not familiar with the cautions, please do follow instructions in this manual for personal safety •...
  • Page 35: Trace Measurement- Auto

    OCC-CWDM-301-18 4.2.3 Trace Measurement- Auto Auto measurement can be applied in case that the length of optic fiber is unidentifiable. OTDR auto select adequate range for measurement. Steps for Auto measurement: • Parameter configuration: for detailed operations, please refer to 3.4.3.2, Parameter Configuration on OTDR Menu Bar.
  • Page 36: Trace Measurement - Manual

    OCC-CWDM-301-18 Note After a certain period of time, the trace displays on the GUI. The trace in the Figure below is a trace during measurement, which will be refreshed for every certain period of time to demonstrate the whole process to users in real time.
  • Page 37: Information Window

    OCC-CWDM-301-18 4.3 Information Window Items of information window: measurement parameters, analysis parameters, and information regarding marker A/B. For details regarding information window, please refer to 3.4.2 OTDR information window. 4.3.1 Switch between Information Window Items Under GUI of Figure 4-4., press [ ] and the items of information window will display in circulation: measurement parameter→...
  • Page 38: Zoom Out Trace Horizontally

    OCC-CWDM-301-18 4.5 Zoom out Trace Horizontally This function is to zoom out trace horizontally. Under GUI of Figure 4-4., use [▲] and [▼] to highlight, and press [Enter] to zoom out trace. or to zoom out trace horizontally by the combining hot key [Shift/ ]+[◄] in the keyboard. 4.6 Zoom in Trace Vertically This function is designed for users to review event points more carefully.
  • Page 39: Save Trace

    OCC-CWDM-301-18 4.9 Save Trace When auto or manual measurement is finished, the measurement trace can be saved. Contents of trace saved include: Trace curve, related information of trace. • Under GUI of Figure 4-4., use [▲] and [▼] to highlight , then press [Enter] to enter, as shown in Figure 4-5.
  • Page 40: Browse Saved Traces

    OCC-CWDM-301-18 4.10 Browse Saved Traces Under GUI of Figure 4-4, use [▲] and [▼] to highlight, press [Enter] to confirm, as shown in Figure 4-6. S e l e c t f i l e : 1 1 8 / 3 0 0 T ra c e001 27- M a y- 2005 10 : 30 T ra c e002...
  • Page 41: Alter Measurement In Realtime Testing

    OCC-CWDM-301-18 Note • Make sure the instrument is power off when connecting to PC through RS232 (or USB) data cable; Make sure it’s fastened, then power off. • USB sustains the electrification hot plug. However USB operation rules must be obeyed while connecting to PC.
  • Page 42: Maintenance And Calibration

    OCC-CWDM-301-18 5. Maintenance and Calibration 5.1 Maintenance and Replacing of Batteries Battery for this instrument is rechargeable NiMH battery. Note Notes for maintenance of batteries in the instrument: • In order for the OTDR (including the batteries) to meet specifications, the storage temperature should be within 15℃...
  • Page 43: Calibration Requirements

    OCC-CWDM-301-18 • For maintenance, always go to qualified or certified professionals Tools for Cleaning Interfaces and Connectors • Optic fiber cleaner (for cleaning of optic connectors) • Optic fiber cleaning rod (for cleaning of optic outputs) • Optic fiber cleaning tissue (for cleaning optic interfaces) •...
  • Page 44: Warranty Information

    OCC-CWDM-301-18 6. Warranty Information Terms of Warranty OTDR is warranted against defective material and workmanship for a period of one (1) year from the date of shipment to the original customer. Any product found to be defective within the warranty period would be repaired or replaced by the Provider free of charge.
  • Page 45 OCC-CWDM-301-18 www.PrecisionRatedOptics.com | 888-545-1254...
  • Page 46 Precision Rated Optics, Inc Corporate Office Billing & Processing PO Box 877 Trexlertown, PA 18087 Precision Rated Optics, Inc Product Distribution Center Manufacturing & Testing 9999 Hamilton Blvd Breinigsville, PA 18031...