13
LAN Configuration
Automatic configuration with Bootp
6 Click OK.
7 Leave Edit BootP Addresses by pressing Close.
8 Exit BootP Settings by pressing OK.
9 After each modification of the BootP settings (i.e. EditBootPSettings) a
10 Power cycle the Instrument.
11 Use the PING utility to verify connectivity by opening a command
278
• Subnet mask
• Gateway address (optional)
The configuration information entered is saved in the Tab File.
stop or start of the BootP service is required for the BootP service to
accept changes. See
or
"Restarting the Agilent BootP Service"
OR
If you changed the IP address, power cycle the instrument for the
changes to take effect.
window and typing:
Ping 192.168.254.11 for example.
The Tab File is located at
C:\Documents and Settings\All Users\Application Data\Agilent\BootP\TabFile
"Stopping the Agilent BootP Service"
1290 Infinity Quaternary Pump User Manual
on page 279
on page 280.