Using the Scaling Function
(Measuring a temperature)
On channel 1 of slot 0 (0001), measure the input signal
ranging from 0 to 5 VDC as 0.0 to 600.0 °C.
Path
MENU key > Browse tab > Setting >
Setting menu > AI channel settings > Range
1
Tap First-CH > 0001.
2
Check that Last-CH is 0001.
3
Tap Type > Volt.
4
Tap Range > 6V.
5
Tap Span Lower, and enter 0.000.
6
Tap Span Upper, and enter 5.000.
7
Tap Calculation > Linear scaling.
8
Drag the screen up.
Show the setting parameters off the screen at the
bottom.
9
Tap Decimal place > 1.
10
Tap Scale Lower, and enter 0.0.
11
Tap Scale Upper, and enter 600.0.
12
Tap Unit > , and enter °C.
13
Tap Save.
44
Power Specialties, Inc. | www.powerspecialties.com | (816) 353-6550
You can register displays that you use frequently as
favorite screens and display them with easy operation.
You can register up to 20 displays.
Registering a Favorite Screen
1
2
1
2
3
4
5
6
7
8
9
10
11
12
3
4
13
5
6
Deleting a Favorite Screen
1
2
3
4
Operation complete
Registering and Deleting Favorite
Screens
Show the display that you want to register as a
favorite screen.
Press MENU.
The menu screen appears.
Add a favorite screen
Favorite screen list
Scroll
Tap Add favorite.
A confirmation screen appears.
Tap Favorite name, and enter the name.
Tap OK.
The display is registered.
Tap the Close icon.
The screen closes.
Press MENU.
Tap Universal tab > Remove favorite.
Select the screen to delete, and tap OK.
Tap the Close icon.
The screen closes.
Close button
Operation complete
Operation complete