MAINTENANCE
(6) If replacing sensor filter at this time:
Carefully peel off old filter taking care not to damage the inside of
the case.
On new filter peel off backing exposing adhesive. Note proper
orientation of filter and apply to inside of front case adhesive side
against case.
Press filter into place taking care not to damage filter surface.
(7) Reinstall the sensor gasket in the case front.
Ensure sensor gasket is properly installed.
(8) Re-install the screws.
(9) Go into the sensor setup and turn ON the sensor.
(10) Verify calibration response and re-calibrate.
Allow sensors to stabilize at least 30 minutes before calibration.
Calibration is required after a sensor is installed; otherwise, the
instrument will not perform as expected and persons relying on this
product for their safety could sustain serious personal injury or death.
4.5.

Cleaning the Instrument

Clean the exterior of the instrument regularly using only a damp cloth.
Do not use cleaning agents.
4.6.

Storage

When not in use, store the instrument in a safe, dry place between 65°F
and 86°F (18 °C and 30 °C). After storage, always recheck instrument
calibration before use.
4.7.

Shipment

Pack the instrument in its original shipping container with suitable padding.
If the original container is unavailable, an equivalent container may be
substituted.
48