Using the VPAP Adapt SV

WARNING
!
Before starting therapy with a new mask select the correct mask type in the
Settings menu (page 22). When adding or removing a new component such
as a mask, humidifier, or anti-bacterial filter, perform the Air Circuit Learn
function (page 22).

Starting Treatment

When the system is fully assembled press the power switch at the back of the
VPAP Adapt SV to ON. The Welcome screen will be displayed while the flow
generator is warming up. Allow the unit to warm up for approximately
30 seconds before placing the mask on your face.
WELCOME
Figure 11: Welcome screen
Once the VPAP Adapt SV has warmed up, breathing into the mask should start
therapy immediately. If not, press the Start/Stop key once to commence
treatment.
The VPAP Adapt SV restarts in the mode in which it stopped.

Stopping Treatment

If SmartStop is enabled treatment will stop automatically when you remove your
mask. If the SmartStop function has been disabled, press the Start/Stop key.

Standby Mode

VPAP ADAPT SV
menu
Figure 12: Standby screen
Once you have stopped treatment the VPAP Adapt SV will be in standby mode.
In standby you can:
press the power switch at the back to OFF to turn off the flow generator
commence treatment again by breathing into the mask or pressing the
Start/Stop key
enter any of the menus by pressing the Left key.
When in standby mode, a gentle (non-therapeutic) flow of air will still be present.
Note: If the SMART DATA - AUTO APPEAR option is set to ON the screens
available under the Results menu will be displayed automatically for
approximately 20 minutes after your session is completed.