Introduction VPAP™ Adapt SV The VPAP Adapt SV flow generator has been designed to be reliable, comfortable and easy to use. You can observe your treatment and make adjustments to settings using a keypad and LCD screen. CAUTION USA federal law restricts this device to sale by or on the order of a physician.
Page 4
The VPAP Adapt SV should be used with masks and accessories • recommended by ResMed or the prescribing physician. Use of incorrect masks and accessories may adversely affect the function of the VPAP Adapt SV. The VPAP Adapt SV is designed for use with masks that allow exhaled gases •...
The following side effects may arise during a course of therapy with the VPAP Adapt SV: drying of the nose, mouth, or throat • bloating • ear or sinus discomfort • eye irritation • skin rashes • chest discomfort. • Quick Setup Guide Attach the air tubing and pressure sensor tube to the VPAP Adapt SV and mask (page 9).
Remove your mask to stop treatment. If SmartStop is enabled, treatment will stop automatically. If the SmartStop function has been disabled, press the Start/Stop key once. VPAP Adapt SV System VPAP Adapt SV Components The VPAP Adapt SV flow generator is supplied with air tubing and a power cord. Masks are supplied separately, as you and your clinician have to decide which mask is best for you.
Page 7
(hose, sensor line, clips, proximal cuff) PN26923. Masks You will also need a ResMed mask system (supplied separately). For the latest available masks, see www.resmed.com. Please refer to the table on page 21 to set up your mask correctly for use with the VPAP Adapt SV. Your clinician can explain the features of the masks available and will discuss the mask that best suits your needs.
There are two sockets at the rear of the flow generator—the top one for an AC (standard mains electricity) and the lower one for a DC power cord (see “Using DC Power to Run the VPAP Adapt SV” on page 15). ResMed recommends using the AC power cord supplied with the unit.
Page 9
Connect air tubing Connect the air tube firmly onto the air outlet at the front of the unit. Figure 4: Front view of VPAP Adapt SV To attach the pressure sensor tube to the unit, screw the Luer connector onto the socket on the side of the unit.
Humidifiers A humidifier may be required if you are experiencing dryness of the nose, throat, or mouth. The following ResMed humidifiers are compatible with the VPAP Adapt SV: HumidAire heated humidifier Humidifier accessory (HumidAire only) Figure 7: Humidifiers compatible with the VPAP Adapt SV When using a humidifier, stop the VPAP Adapt SV by using the Start/Stop key or by removing the mask to trigger the SmartStop function.
Page 11
HumidAire 2i™ The HumidAire 2i provides heated humidification for the VPAP Adapt SV. To attach the HumidAire 2i, remove the VPAP Adapt SV front cover and attach the humidifier docking station and water chamber. No other accessories are required for use. Please refer to the HumidAire 2i User’s Manual for details. Once you have performed the Air Circuit Learn function the VPAP Adapt SV is ready for use.
Page 12
HumidAire™ You will need a medium size 21" (52 cm) air tube to connect the VPAP Adapt SV unit to the HumidAire™ (see Figure 7). Make sure both the HumidAire and the VPAP Adapt SV are turned off. Fill the HumidAire with water as described in the humidifier manual. Place the filled water chamber inside the HumidAire.
Page 13
Connect the mask system to the free end of the long air tubing. The final assembly should look like this: Switch on the HumidAire and the VPAP Adapt SV. WARNING Make sure that the power cords and plugs are in good condition and the equipment is not damaged.
ResLink™ The ResLink is a device that records data onto a SmartMedia™ card during your therapy. The card can be returned to your clinician so that they can observe your progress and adjust flow generator parameters if necessary. The ResLink connects to the back of VPAP Adapt SV as shown.
Using DC Power to Run the VPAP Adapt SV The VPAP Adapt SV can be run on DC in case of power failure. ResMed supplies separately a converter that allows a 12V or 24V battery to be connected to the DC input of the VPAP Adapt SV.
Using the VPAP Adapt SV WARNING Before starting therapy with a new mask, select the correct mask type in the Settings menu (page 21). When adding or removing a new component, such as a mask, humidifier, or anti-bacterial filter, perform the Air Circuit Learn function (page 21).
Control Panel Functions The control panel of the VPAP Adapt SV has a keypad that allows you to: start or stop treatment • adjust settings on the flow generator • mute or stop alarms • view treatment data. • QuickView key Start/Stop Figure 13: The VPAP Adapt SV control panel LCD screen...
Left key (green) Enters a menu. • Confirms and applies settings. • Performs the function indicated by the text above it in the LCD (eg, ’enter’, • ’change’). Right key (red) Performs the function indicated by the text above it in the LCD (eg, ’exit’). •...
Page 19
Example You wish to change the setting for the mask you are using. VPAP Adapt SV menu SETTINGS enter exit MASK: FULL FACE change LEARN CIRCUIT? SMARTSTOP: OFF change SOUND LEVEL: LOW change RESULTS enter exit OPTIONS enter exit SERVICING enter exit Figure 14: Using the keys to adjust settings...
Menus The options available in the Settings menu will depend on the therapy you are receiving (either ASV or CPAP), as set by your clinician. Figure 15: Navigating the Patient menu. VPAP ADAPT SV menu SETTINGS enter exit ASV MODE MASK: FULL FACE change LEARN CIRCUIT?
Settings Menu You will only see the options relevant to your therapy. In CPAP (Continuous Positive Airway Pressure) mode, the clinician sets a pressure that will be constant throughout treatment. In ASV mode, the VPAP Adapt SV continually measures your breathing and regulates the air pressure accordingly. Mask Select the type of mask that you will be using (you will only need to adjust this if you have more than one mask).
7 . When the function is complete, a screen will appear saying that the circuit has been learned successfully. The next menu item is then displayed. Commence treatment as described in “Starting Treatment” on page 16. If there is a problem with the circuit a screen saying ’Invalid circuit’ will be displayed instead of the progress screen.
Mask-fit Gives a rating of how well the mask was fitting during your last session on the VPAP Adapt SV. The stars indicate how good the fit was (more stars = better fit). Star rating ***** ****- ***-- **--- *---- HIGH LEAK Average Pressure Provides information on the average air pressure during your most recent...
Servicing Menu This menu displays the following information: SN—the flow generator’s serial number • PCB—the flow generator’s PCB number (PCB—printed circuit board) • SW—the version of software currently installed on the flow generator. • Treatment Screens There are three Treatment screens displayed in the VPAP Adapt SV LCD. The LCD will display the Treatment screens when treatment commences, and during treatment if no key is pressed for 20 minutes.
1. Check that the air tube and sensor tube are connected properly. 2. Turn the VPAP Adapt SV off and on again at the power switch. If the alarm persists, return the unit to ResMed for servicing. Action Check tube connections.
6. Try using the flow generator one more time. 7 . If the high pressure alarm activates repeatedly, discontinue use and return to ResMed for servicing. If the alarm does not recur, then continue to use as normal. Note: The alarm system is designed to ignore coughing. However, if a cough is especially intense and prolonged, it may trigger the high pressure alarm.
3. Check whether there is a blockage in the air circuit. 4. Remove blockage. 5. Turn power back on. 6. If the flow blocked alarm activates repeatedly, discontinue use and return to ResMed for servicing. If the alarm does not recur, then continue to use as normal. Alarms...
Will stop when: blockage is removed. • Cleaning and Maintenance You should regularly carry out the cleaning and maintenance described in this manual. CAUTION Do not wash the pressure sensor tube. If fluid enters the pressure sensor tube allow it to dry completely by hanging it in a clean place out of direct sunlight.
Air tubing Remove the pressure sensor tube from air tubing Figure 19: Air tubing assembly removed for cleaning CAUTION Do not use bleach, chlorine-, alcohol-, or aromatic-based solutions • (including all scented oils), moisturising or antibacterial soaps to clean the cushion, mask, air tubing, or the VPAP Adapt SV. These solutions may cause hardening and reduce the life of the product.
Hypoallergenic Air Filter The ResMed hypo-allergenic filter is a standard filter to which an electrostatic filter has been bonded to capture very small particles. The filter is intended for those users of ResMed flow generators who will benefit from enhanced filtering of the air delivered during treatment.
Applicable ResMed warranty details are provided with the device at the time of original supply. Of course, as with all electrical devices, if any irregularity becomes apparent, you should exercise caution and have the device inspected by an authorized ResMed service center.
Problem No LCD display FAULT message displayed on LCD System Error 7 message Glossary Alarms When more than one alarm is activated at the one time, the highest priority alarm will be displayed. Backlight The backlight turns off if no button has been pressed for two minutes. It can also be kept permanently on by selecting On in the Backlight screen of the Options menu.
SmartStop A feature that stops the VPAP Adapt SV automatically when you remove the mask. You can turn SmartStop on or off in the Settings menu. Tidal volume The volume of air, in liters, breathed in or out in one breath. Technical Specifications Dimensions (H x W x D) 5.6"...
Page 34
Alarm specifications High maximum mask pressure alarm • Stops air delivery at 30 cm H High mean mask pressure alarm • Stops air delivery when mean mask pressure exceeds 22 cm H over approx 1 minute). Low mask pressure alarm •...
Displayed values Value Range Proximal cuff Pressure -3 to 40 cm H Mass flow sensor Leak 0–99 liters per minute Respiratory rate 8–30 breaths per minute Tidal volume 0–4500 milliliters Minute ventilation 0–40 liters per minute Target MV 0–40 liters per minute Maximum low impedance flow at stated pressures Pressure (cm H 13.0...
Guidance and Manufacturer’s Declaration — Electromagnetic Emissions and Immunity Guidance and manufacturer’s declaration – electromagnetic emissions The VPAP Adapt SV is intended for use in the electromagnetic environment specified below. The customer or the user of the VPAP Adapt SV should assure that it is used in such an environment. Emissions test RF emissions CISPR 11 RF emissions CISPR 11...
Page 37
Guidance and manufacturer’s declaration – electromagnetic immunity The VPAP Adapt SV is intended for use in the electromagnetic environment specified below. The customer or the user of the VPAP Adapt SV should assure that it is used in such an environment. IEC60601-1-2 Immunity test test level...
Page 38
Guidance and manufacturer’s declaration – electromagnetic immunity (continued) The VPAP Adapt SV is intended for use in the electromagnetic environment specified below. The customer or the user of the VPAP Adapt SV should assure that it is used in such an environment. Immunity IEC60601-1-2 test...
Page 39
Recommended separation distances between portable and mobile RF communications equipment and the VPAP Adapt SV The VPAP Adapt SV is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the VPAP Adapt SV can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the VPAP Adapt SV as recommended below, according to the maximum output power of the communications equipment.
ResMed shall not be responsible for any incidental or consequential damages claimed to have occurred as a result of the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you.
Index air circuit learn function 21 air filter 30 hypoallergenic 30 air pressure monitoring 23 air tubing attaching to mask 9 cleaning 28 connecting 9 air tubing system 6 alarm flow blocked 27 high leak 27 high pressure 26 low mask pressure 25 power fail 26 alarm LED 18 anti-bacterial filter 14...
Need help?
Do you have a question about the Flow Generator VPAP Adapt SV and is the answer not in the manual?
Questions and answers