Power requirements
Site Preparation Guide
Table 2
Power configurations
Configuration
Single phase,
120 VAC
Single phase,
200-240 VAC
208-Wye
* Varies with country and/or region
Table 3 lists the power requirements for the 5973 Series MSD
and related equipment. Extra power capacity for future
additions is a very good idea.
Each product listed requires a dedicated circuit. The MSD, GC,
and data system must each have a separate circuit breaker. All
of the equipment must share a common ground.
Power must meet the stability specifications listed in Table 3.
Use a line monitor to check power stability. If your line power is
unstable, you may need to install a line conditioner.
Measurement
Line to neutral
Line to ground
Ground to neutral
Line to neutral
Line to ground
Ground to neutral
Line to line
(phase A to phase B)
Line to ground
(phase A to ground)
Line to ground
(phase B to ground)
Site Preparation
Nominal voltage
*
120 VAC
120 VAC*
< 0.5 V rms
200, 220, or 240 VAC*
200, 220, or 240 VAC*
< 0.5 V rms
208 VAC
120 VAC
120 VAC
25