10 Troubleshooting

WARNING
Read all general guidelines, warnings and cautions before performing any repair/maintenance on the bed. All repair and
maintenance on the bed should be performed by a qualified technician.
Unplug the power cord from its power source before performing any maintenance on the electric bed. DO NOT unplug
the power cord from the control box. Damage to cord will result.
EFFECT
Bed does not stay in
Floor lock not engaged.
place
Insufficient static friction
between floor lock foot
bumper and floor.
Floor lock not functioning.
Bed is not level
Hi/Lo actuators are unevenly
extended
Actuator not working.
System Fault
Needs Reset
Part No 1153410
POSSIBLE CAUSE
Green plunger on floor lock post is
visible.
Floor surface is slippery and/or
foreign membrane between bumper
and floor.
Floor lock is stuck in one position.
Bed deck is not parallel with floor
Flashing LEDs on ACP, Bed function
will not operate. The control box
may beep when pendant function is
pressed.
VERIFICATION
Activate floor lock - green plunger should NOT be
visible.
Clean the surrounding floor and remove loose
material under the bed.
Contact Invacare Continuing Care Service
Department 1-800-668-2337.
Raise bed to highest position
or
Lower bed to lowest position
To Reset - hold the pendant bed up and bed down
buttons for 5+ seconds. The bed will beep and stop
beeping when reset is complete. Then press the bed
up button. If the bed beeps, the actuators are out of
sync. Continue rising to the highest position to
synchronize the actuators.
65
10 TROUBLESHOOTING
CORRECTIVE ACTION
Invacare ®CS Series Beds