Message Codes

Me s s a g e Co d e s
No.
Cause
156
Transverse tilt greater than 10°
162
Calibration error
204
Calculation error
240
Data transfer error
252
Temperature too high
253
Temperature too low
255
Received signal too weak, measuring time too long
256
Received signal too high
257
Too much background light
260
Laser beam interrupted
301
Device was moved, levelling not valid anymore
303
Error with Leica DST 360 adapter
* If other message codes are displayed frequently even the instrument has been switched off and on, please contact your dealer.
Leica DISTO™ X4
Correction
Hold the instrument without any transverse tilt.
Make sure the device is placed on an absolutely horizontal and flat
Perform measurement again.
Connect device and repeat procedure
Let device cool down.
Warm device up.
Change target surface (e.g. white paper).
Change target surface (e.g. white paper) .
Shadow target area.
Repeat measurement.
Perform levelling again. Measuring with invalid levelling is partially
Repeat measurement.