Troubleshooting; Display Board Checks; Table 4-2 Display Board Checks - Keithley 2010 Service Manual

Multimeter
Hide thumbs Also See for 2010:
Table of Contents

Advertisement

4-10

Troubleshooting

Troubleshooting
Troubleshooting information for the various circuits is summarized below. See Analog signal
switching states for additional information regarding the analog circuitry.

Display board checks

If the front panel DISP test indicates that there is a problem on the display board, use
Table
4-2. See Principles of operation for display circuit theory.
Table 4-2
Display board checks
Step
Item/component
1
Front panel DISP test.
2
P1005, pin 5
3
P1005, pin 9
4
U401, pin 1
5
U401, pin 43
6
U401, pin 32
7
U401, pin 33
Required condition
Verify that all segments operate.
+5V ±5%
+37V ±5%
Goes low briefly on power up, then
goes high.
4MHz square wave.
Pulse train every 1ms.
Brief pulse train when front panel key
is pressed.
Remarks
Use front panel display test.
Digital +5V supply.
Display +37V supply.
Microcontroller RESET.
Controller 4MHz clock.
Control from main processor.
Key down data sent to main processor.