3 Specifications

3.1 Specifications

Table 2 Specifications

Specifications
Temperature range
Nominal resistance at 0.01 °C
Temperature coefficient
Accuracy
[1]
Short-term repeatability
Drift
[3]
Hysteresis
Sheath length
Sheath diameter
Sheath material
Transition junction temperature
range
[4]
Transition junction dimensions
Sensor length
Sensor location
Minimum insulation resistance
Minimum immersion length
Maximum immersion depth in
liquid medium
Maximum immersion depth in
dry medium
Response time
[5]
Self heating (in 0 °C bath)
Lead-wire cable type
Lead-wire length
Lead-wire temperature range
Calibration
Includes calibration and 100 hr drift (k = 2).
[1]
Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k = 2).
[2]
After 100 hours at max temp, 95 % confidence (k = 2).
[3]
Temperatures outside this range will cause irreparable damage.
[4]
Per ASTM E 644.
[5]
5606
–200 °C to 160 °C
± 0.04 °C at 0.01 °C
± 0.06 °C at 160 °C
± 0.03 °C at 0.01 °C
[2]
± 0.04 °C at 160 °C
± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C
50 mm ± 5 mm
(2 in ± 0.2 in)
316 SST
-200 °C to 160 °C
No transition
20 MW at 23 °C
Full immersion
Full immersion
Full immersion 
12 seconds typical
Enameled copper wire
160 °C
Calibration not included. NVLAP accredited calibration available. See
ordering information.
100 W ± 0.1 W
0.00385 W/W/°C
± 0.020 °C at 0.01 °C
± 0.035 °C at 450 °C
± 0.015 °C
3.1 mm ± 0.1 mm (1/8 in ± 0.004 in)
30 mm ± 3mm (1.2 in ±0.1 in)
3 mm ± 1 mm from tip (1.2 in ± 0.04 in)
Use SST or glass tube to protect
± 0.003 °C
2.4 meters (8 ft)
Specifications
Specifications
5607
0 °C to 450 °C
± 0.04 °C at 0.01 °C
± 0.07 °C at 450 °C
± 0.02 °C at 0.01 °C
± 0.04 °C at 450 °C
40 mm ± 5 mm
(1.6 in ± 0.2 in)
Inconel
600
®
0 °C to 450 °C
35 mm x 7 mm
(1.4 in x 0.3 in)
100 MW at 23 °C
Full immersion
transition junction
Full immersion
9 seconds typical
SST over-braid
450 °C
5